Electromagnetic Environment (Continued)
Alaris
System (Continued)
®
Reduce the potential for electromagnetic interference by maintaining a minimum distance between
portable and mobile
recommended in this table, based on the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed in this table, the recommended separation
distance (d) in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where
P
transmitter manufacturer.
Rated Maximum
Output Power of
Transmitter (W)
0.01
0.1
1
10
100
A-10
Regulations and Standards
Compliance (Continued)
Recommended Separation Distances
communications equipment (transmitters) and the Alaris
RF
is the maximum output power rating of the transmitter in watts (
Separation Distance Based on Transmitter Frequency (m)
k
z -
z
150
H
80 MH
Outside
ISM
Bands
3.5
— d = [-----] √P
V
1
0.35
0.12
0.35
1.11
3.5
Table 4
kHz -
z
150
80 MH
In
Bands
ISM
12
— d = [-----] √P
V
2
0.12
0.38
1.2
3.8
12
System as
®
W
z -
z
80 MH
800 MH
12
— d = [-----] √P
E
1
0.12
0.38
1.2
3.8
12
Alaris
System Directions for Use
®
Appendix Section
) based on the
z -
z
800 MH
2.5 GH
23
— d = [-----] √P
E
1
0.23
0.73
2.3
7.3
23