Reference the PC Unit and module-specific Sections of this
for the following references:
DFU

Alarms, Errors, Messages

Audio Characteristics
Definitions
Radio Frequency Note
The CI Board has self-testing capability, both at power on and
during normal operation. If communication with the CI Board
fails, a visual message displays to indicate a communication
error. A communication error to an outside access point does
not cause a Alaris
System alarm or visual message.
®
Message
NETWORK COMM ERROR
8-6
Troubleshooting and Maintenance
Alarms, Errors, Messages

Messages

Communication error occurred
between C I Board and Alaris
Server.
Meaning
®
Response
Power Alaris
System off and then
®
back on. If error is repeated:
To continue manual-only
operation, press
CONFIRM
key.
Replace PC Unit as soon as
possible.
Refer instrument to be serviced
by qualified personnel, to:
- reprogram/reconfigure
C I Board
- verify wireless boot version
- replace C I Board, as
necessary
Alaris
System Directions for Use
®
CI Board Accessory Section