Features and Definitions (Continued)
Pump Module
Auto-Restart
Default Occlusion Pressure
Free Flow Protection
KVO Rate Adjust
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Features and Displays (Continued)
Part of Alaris
designed to minimize nuisance, patient-side occlusion alarms.
Allows system to automatically continue an infusion following
detection of a patient-side occlusion if downstream pressure falls to
an acceptable level within a
this feature is enabled, "Checking Line" function occurs when
downstream pressure exceeds pressure limit.
• In Selectable Pressure Mode: Pressure limit will be either user
adjustable or "locked" in system configuration.
• In Pump Pressure Mode: Pressure limit is a function of flow rate
and is automatically determined by device.
If downstream pressure decreases to a predetermined level, (below
of pressure limit) during
50%
infusion automatically continues. If condition is not cleared within
seconds, a "Partial Occlusion - Patient Side" alarm occurs.
15
Using Guardrails
(zero) to
0
allowable number of restarts is exceeded or if feature is set to zero,
an "Occluded - Patient Side" alarm occurs when system detects
downstream pressure that exceeds pressure limit.
Starting occlusion pressure limit which can be configured by profile in
mmHg increments.
25
All Alaris
utilize a unique clamping device (safety clamp on pumping
mechanism) to prevent inadvertent free-flow when administration set
is removed from instrument.
Used to select
which is rate of fluid flow after an "Infusion Complete" occurs.
rate will never exceed infusion rate.
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
System's Downstream Occlusion Detection system
®
15
Editor Software, system can be configured to allow
®
restart attempts within a rolling
9
Pump Module/Gemini Infusion System administration sets
®
(Keep Vein Open) rate (
KVO
-second "Checking Line" period. If
-second "Checking Line" period,
15
minute period. If
10
to
0.1
General Information
mL/h allowed),
20
KVO
2-83