Primary Infusion
IV Fluid Infusion (Continued)
Volume/Duration Infusion
1.
Press
VOLUME DURATION
2.
To enter
(Syringe Module: instead of infusing
VTBI
press
soft key and use numeric data entry keys.
VTBI
OR
Syringe Module: To deliver entire contents of syringe,
leave
as
VTBI
ALL
3.
To enter volume duration, press
use numeric data entry keys.
• Rate is automatically calculated.
4.
Verify correct infusion parameter entry and press
soft key.
• If programmed
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
If programmed
•
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
- With Guardrails
soft key.
.
DURATION
Fluid is outside Soft Limit for that
IV
Fluid is outside Hard Limit for that
IV
Suite MX Protection (Continued)
®
)
ALL
,
soft key and
START
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Fluid Setup
A
TPN
PRIMARY INFUSION
4.17
RATE =
mL/h
ALL
VTBI
12 00
DURATION
h
min
>Select DURATION
DELAY
RATE
START
OPTIONS
VOLUME
Programming
2-37