Setting Loading Dose Only (Continued)
Setting Loading Dose
from Infusion Mode Screen (Continued)
7.
Press
CONFIRM
• Upon pressing Channel Select on PCA Module,
Infusion Mode screen becomes available for selection
of infusion mode.
NOTE:
Loading dose is included in the
Max Limit.
The following procedures should be used only when setting a
using the Drug Library.
BOLUS DOSE
1.
Press
CHANNEL SELECT
2.
Press
BOLUS DOSE
3-20
Programming
Infusion Modes (Continued)
soft key.
but is not included in the
VTBI
Setting Bolus Dose
.
soft key.
A
The loading dose has
completed.
>Press CONFIRM
Summary
A
Morphine
Cont. Dose
Max Limit:
[Conc]: 1 mg/mL
>Press START to Close
Summary
PRO-
GRAM
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
CONFIRM
2
mg/h
30
mg/4 h
BOLUS
START
DOSE