Setting Bolus Dose - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Setting Loading Dose Only (Continued)
Setting Loading Dose
from Infusion Mode Screen (Continued)
7.
Press
CONFIRM
• Upon pressing Channel Select on PCA Module,
Infusion Mode screen becomes available for selection
of infusion mode.
NOTE:
Loading dose is included in the
Max Limit.
The following procedures should be used only when setting a
using the Drug Library.
BOLUS DOSE
1.
Press
CHANNEL SELECT
2.
Press
BOLUS DOSE
3-20
Programming
Infusion Modes (Continued)
soft key.
but is not included in the
VTBI

Setting Bolus Dose

.
soft key.
A
The loading dose has
completed.
>Press CONFIRM
Summary
A
Morphine
Cont. Dose
Max Limit:
[Conc]: 1 mg/mL
>Press START to Close
Summary
PRO-
GRAM
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
CONFIRM
2
mg/h
30
mg/4 h
BOLUS
START
DOSE