Directions for Use ® Alaris System (with PC Unit, Model 8000) Supports Guardrails ® Suite MX with Guardrails ® Point-Of-Care software and v9 Operating System software. February 2007 % SpO 2 RATE (mL/h) RATE (mL/h) RATE (mL/h) PULSE (BPM) CHANNEL...
General Contact Information Cardinal Health ® Alaris Products Wateridge Circle 10221 San Diego, California 92121 cardinal.com/alaris Customer Advocacy - North America Clinical and technical feedback. Phone: 800.854.7128, Ext. 7812 E-Mail: CustomerFeedback@cardinal.com Technical Support - North America Maintenance and service information support; troubleshooting.
A Data Set is developed and approved by the facility’s own multi-disciplinary team using the Editor Software, the based ® authoring tool. A Data Set is then transferred to the Alaris System by qualified personnel. The approved Data Sets are maintained by the Editor Software for future updates and reference.
Product-specific warnings and cautions, covered in the applicable Sections of this DFU provide information needed to ® safely and effectively use the Alaris System. DANGER is an alert to an imminent hazard which could result in serious personal injury and/or product damage if proper procedures are not followed.
Page 6
Instruments are tested and calibrated before they are packaged for shipment. To ensure proper operation after shipment, it is recommended that an incoming inspection be performed before placing the instrument in use. Prior to placing the Alaris ® System in use: Perform check-in procedure using System Maintenance software.
Page 7
® Alaris PC Unit Model 8000 SILENCE SYSTEM OPTIONS ENTER CLEAR CANCEL ® Alaris System (with v9 Model 8000) DFU, Section...
Page 9
..................................Operating Features, Controls, Indicators 1-35 ............................Displays 1-37 ........................................SYSTEM CONFIGURABLE SETTINGS 1-38 ..............................SPECIFICATIONS AND SYMBOLS 1-39 ................................Specifications 1-39 ......................................Symbols 1-40 ........................................Alaris ® System (with v9 Model 8000) DFU Table of Contents PC Unit Section...
Page 10
1-47 ..............................Battery Type and Charging 1-47 .................................. Battery Charge 1-48 ......................................Battery Care 1-48 ....................................... Battery Cautions and Disposal 1-48 ................................INSPECTION REQUIREMENTS 1-49 ................................. ® Alaris System (with v9 Model 8000) DFU Table of Contents PC Unit Section...
• System Maintenance software (and its instructions) for Alaris System check-in, maintenance, and wireless configuration CAUTION The PC Unit is the core of the Alaris ® System and provides a common user interface for programming infusions and monitoring, which helps to reduce complexity at the point of care.
Page 12
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K Getting Started Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Application of adhesive tape or other materials to the sides of the PC Unit and modules may prevent proper latching. ® The Alaris System is designed to operate a maximum of infusion or monitoring modules. Modules added in excess of are recognized by the system.
Push module release latch and then rotate module(s) up and away from PC Unit or attached module (opposite to motion shown above) to disengage connectors. • Alaris ® System reidentifies and shows appropriate module identification ( ), from left to right.
If the affected module operates normally when it is attached via an alternate connector, it may be used until a replacement module can be substituted. Alaris ® System (with v9 Model 8000) DFU General Setup and Operation PC Unit Section...
Module A: YYYY-MM-DD at power up. Remove and, if needed, replace module requiring maintenance with a new module (see "Attaching and Detaching Modules"). If Alaris ® System was powered off to replace PC Unit, CONFIRM reinitiate start-up process. If an attached module (such as a Pump Module) was...
Page 18
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Auto-ID Module Section of this DFU. >Enter Patient ID and Press CONFIRM PAGE EXIT CONFIRM DOWN General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
To return to PC Unit screen, press soft key. MAIN SCREEN • After seconds without a key press, Main Display Softer Louder appears. >Change Setting or Cancel MAIN SCREEN Alaris ® System (with v9 Model 8000) DFU General Setup and Operation PC Unit Section...
An advisory tone (if Key Click Audio is enabled) and a three-second prompt on Main PANEL UNLOCKED Display confirm activation. PANEL UNLOCKED VOLUME AUDIO INFUSED ADJUST 1-10 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Power Down All Channels Anesthesia Mode >Select an Option or EXIT PAGE EXIT DOWN Adjust display and return to main screen (see "Start-Up", "Adjusting Display Contrast" procedure). Alaris ® System (with v9 Model 8000) DFU 1-11 General Setup and Operation PC Unit Section...
To verify correct entry, press soft key. CONFIRM Patient ID Entry 123456789CD_ _ _ _ _ >Enter Patient ID and Press CONFIRM PAGE EXIT CONFIRM DOWN 1-12 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Page 23
Patient ID Entry • New Patient ID Entry verification screen appears. 234567891EF_ _ _ _ _ >Enter Patient ID and Press CONFIRM PAGE EXIT CONFIRM DOWN Alaris ® System (with v9 Model 8000) DFU 1-13 General Setup and Operation PC Unit Section...
System Options 1 of 3 Display Contrast Patient ID Clinician ID Power Down All Channels Anesthesia Mode >Select an Option or EXIT PAGE EXIT DOWN 1-14 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
System Options 1 of 3 Display Contrast Patient ID Clinician ID Power Down All Channels Anesthesia Mode >Select an Option or EXIT PAGE EXIT DOWN Alaris ® System (with v9 Model 8000) DFU 1-15 General Setup and Operation PC Unit Section...
All limits are set to Soft. patient will be taken to when the Anesthesia Mode is discontinued. • Dose checking mode is set to Smart. This ensures that the Alaris ® System will be in the correct profile following •...
Page 27
The Anesthesia Mode can be disabled, and normal operation resumed, using either of the following three methods: • System Options menu. • Disconnecting from power. • Connecting to power. Alaris ® System (with v9 Model 8000) DFU 1-17 General Setup and Operation PC Unit Section...
Page 28
Press soft key. CONFIRM Anesthesia mode was discontinued when AC power cord was disconnected. Press CONFIRM to continue normal operation. >Press CONFIRM CONFIRM 1-18 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
2 of 3 Battery Runtime System Configurations Serial Numbers Software Versions Time of Day >Select an Option or EXIT PAGE PAGE UP EXIT DOWN Alaris ® System (with v9 Model 8000) DFU 1-19 General Setup and Operation PC Unit Section...
Page 30
Max Pt. BSA: Max Pt. weight: 500 kg Patient ID Entry: Disabled Pending IV Orders: Enabled >Press CANCEL or EXIT PAGE PAGE EXIT DOWN 1-20 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Module B: nnnn-nnnnnnnn Module C: nnnn-nnnnnnnn Module D: nnnn-nnnnnnnn >Press CANCEL or EXIT EXIT NOTE: "nnnn-nnnnnnnn" in the illustrated display represents a serial number. Alaris ® System (with v9 Model 8000) DFU 1-21 General Setup and Operation PC Unit Section...
Main processor: nn.nn nn.nn Main boot block: Keyboard: nn.nn >Press CANCEL or EXIT EXIT NOTE: "nn.nn" in the illustrated display represents a software version. 1-22 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
>CONFIRM Time-of-Day EXIT CONFIRM Enter current Time of Day. System Options Time of Day Current time: Change Time __:__ >Enter Current Time EXIT CONFIRM Alaris ® System (with v9 Model 8000) DFU 1-23 General Setup and Operation PC Unit Section...
This option is available only when the Communications Interface Board Accessory is installed and the PC Unit has ® been flashed to support wireless communications with Alaris Server or Mobile Systems Manager. The displayed status updates immediately when a status change takes place.
Page 35
Network Status Status : CONNECTED Uptime : 03:45:35 Bytes Sent: 13, 890 Bytes Recv: 1,200,150 >Press CANCEL to Exit SERVER WIRELESS ADDRESS STATUS STATUS Alaris ® System (with v9 Model 8000) DFU 1-25 General Setup and Operation PC Unit Section...
STATUS Wireless Connection This option is available only when the Communications Interface Board Accessory is installed and the PC Unit has ® been flashed to support wireless communications with Alaris Server or Mobile Systems Manager. 1. Press key. OPTIONS Press soft key times.
To comply with regulations and prevent potential interference with aircraft communications, disable To enable wireless communication, press Enable soft key. wireless communication when the ® Alaris System is used in an aircraft. System Options Wireless Enable Connection Disable >Press ENABLE or...
Module A: YYYY-MM-DD Module B: YYYY-MM-DD Module C: YYYY-MM-DD Module D: YYYY-MM-DD >Press CANCEL or EXIT PAGE EXIT DOWN -- Continued Next Page -- 1-28 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Page 39
Bar Code: YYYY-MM-DD (Hand held) >Press CANCEL or EXIT PAGE UP EXIT NOTE: soft key appears only if an Auto-ID Module is PAGE DOWN attached. Alaris ® System (with v9 Model 8000) DFU 1-29 General Setup and Operation PC Unit Section...
Page 40
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 1-30 General Setup and Operation Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Electrical shock hazard. Do not open case. Refer to qualified service personnel. • To comply with regulations and prevent potential interference with aircraft communications, disable wireless communication when the Alaris ® System is used in an aircraft (see "System Options", "Network Status"). ®...
System near Magnetic Resonance Imaging ( ), including Stereotaxis technology. ® • Do not use the Alaris System near Therapeutic Radiation equipment, such as Linear Accelerators. • Use of any accessory, transducer or cable other than those specified may result in increased emissions or decreased Alaris ®...
(data processing equipment) or IEC 60950 (electromedical equipment). IEC 60601–1 ® • The Alaris System is intended for use by healthcare professionals only. This is a Class Group CISPR 11 medical system. In a domestic environment, this system may cause radio interference.
Page 44
Allow system settings to be customized. If Profiles feature is enabled, system settings defined for selected profile are automatically activated. Tamper Resist Provides a quick one-touch lockout of front panel keypad. 1-34 General Information Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Silence Key: When pressed during an alarm, silences audio System On Key: When for 2 minutes. ® pressed, changes Alaris System from Standby to Operating mode. Options Key: When pressed, allows access to available System or Channel Options. Up/Down Arrows:...
Page 46
90° attachment to a bed rail instead of a pole. Primary Audio Speaker Connector Plug over RJ45 Communication Data Port Tamper Resist Switch Option Upgrade Panel 1-36 General Information Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Nonhighlighted indicates a nonselected soft key. "Active" Soft Key Highlighted indicates a selected soft key. >Press START Prompt Bar SECOND- START PAUSE Look here for user prompts. Alaris ® System (with v9 Model 8000) DFU 1-37 General Information PC Unit Section...
System Configurable Settings If the configuration settings need to be changed from the Factory default settings, reference the applicable Technical Service Manual or contact Cardinal Health Technical Support, for technical, troubleshooting, and preventive maintenance information. With the Profiles feature enabled, the settings are configured independently for each profile.
Canadian electrical safety and U.S. performance standards. Communications connector for attachment. RS-232 Consult operating instructions. Type defibrillation-proof equipment. Electrostatic discharge ( 1-40 General Information Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Page 51
Caution: Federal ( ) law restricts this device to sale by or on order of a physician. U.S.A. " " SYSTEM ON Tamper Resist activate/deactivate switch. Alaris ® System (with v9 Model 8000) DFU 1-41 General Information PC Unit Section...
Page 52
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 1-42 General Information Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Service Manual and System Maintenance software. Alarms, Errors, Messages ® To enhance safety and ease of operation, the Alaris System provides a full range of audio and visual alarms, errors, and messages.
SILENCE Audio Characteristics ® The Alaris System provides various types of alert information. See the product-specific Section of this DFU that applies to the attached module(s) for audio characteristics specific to that module. Type Sound Notes...
Replace CONFIRM module with an operational instrument. Defective Battery Defective battery. To continue temporary operation, press key. Replace SILENCE PC Unit with an operational instrument. Alaris ® System (with v9 Model 8000) DFU 1-45 Troubleshooting and Maintenance PC Unit Section...
Last module powering off. System Powering Down Press any key, except SYSTEM shuts off in indicated number of key, to cancel power down seconds. sequence. 1-46 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU PC Unit Section...
Batteries should be charged in a room with a temperature between ° ° ), to minimize charge time and 80.6 maximize battery life. Alaris ® System (with v9 Model 8000) DFU 1-47 Troubleshooting and Maintenance PC Unit Section...
Battery Care The battery capacity should be checked at least once every ® months. Reference the Alaris System Technical Service Manual for test and replacement procedures. If the PC Unit is to be stored at temperatures in excess of °...
Inspection Requirements ® To ensure the Alaris System remains in good operating WARNING condition, both regular and preventive maintenance inspections Failure to perform these are required. Reference the System Maintenance software for inspections may result in improper detailed instructions. instrument operation.
Page 61
® Alaris Pump Module, Model 8100 ® Alaris Syringe Module, Model 8110 RATE (mL/h) RATE (mL/h) CHANNEL SELECT CHANNEL SELECT PAUSE PAUSE CHANNEL CHANNEL RESTART RESTART ® Alaris System (with v9 Model 8000) DFU, Section...
Page 63
MULTIDOSE MODE 2-68 ......................................Volume/Duration Enabled 2-70 ..................................Volume/Duration Disabled 2-71 ................................... SELECTING PRESSURE LIMIT 2-73 .................................. Pump Module 2-73 ......................................Syringe Module 2-74 ......................................® Alaris System (with v9 Model 8000) DFU Table of Contents Pump and Syringe Modules Section...
Page 64
2-116 ..................................... Alarms 2-116 ........................................... Errors 2-121 ..........................................Messages 2-122 ........................................Possible End of Infusion Messages and Alerts 2-124 .......................... INSPECTION REQUIREMENTS 2-124 ................................. Alaris ® System (with v9 Model 8000) DFU Table of Contents Pump and Syringe Modules Section...
Syringe Module Set Compatibility Card • Syringe Module Technical Service Manual • System Maintenance software (and its instructions) for Alaris ® System check-in, maintenance, and wireless configuration The Pump and Syringe Modules are intended for facilities that utilize infusion and/or syringe pumps for the delivery of fluids,...
Page 66
Introduction (Continued) Contraindications: None known. Electromagnetic Environment: See "Appendix" Section of this DFU ("Regulations and Standards", "Compliance"). Getting Started Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
(see "General Information", "Safety Clamp Fitment"). * "Safety clamp" is referred to on the device as "Flo-Stop". Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 68
Pump Module door. Open roller clamp. Verify no fluid is flowing through drip chamber. NOTE: "Safety clamp" is referred to on the device as "Flo-Stop". Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
(Reference set’s directions for use.) Insert administration set spike into prepared fluid container, following accepted hospital/facility procedure, and hang container inches above Pump Module. Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 70
Prepare administration set (reference Set Compatibility Card, provided separately) in accordance with manufacturer’s directions for use. Attach upper fitting of administration set to syringe tip. Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Ensure instrument is as close to level of patient as CHANNEL SELECT possible (patient should be in line with key). RATE (mL/h) Align with Patient CHANNEL SELECT PAUSE CHANNEL RESTART Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 72
Insert syringe (from front of instrument) by sliding flat edge of syringe barrel flange between barrel flange grippers. RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 73
Ensure the syringe plunger head makes contact with the small black sensor, located on the bottom of the drive head (between the plunger grippers). Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 74
CHANNEL • cavity forward (membrane toward instrument) RESTART Gently slide pressure sensing disc up into slot in pressure sensing disc housing. Pressure Sensing Disc 2-10 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Fast Start WARNING Priming - Using Options Menu When priming: The Priming option can be enabled at the time the Alaris ® PRIME System is configured for use. The Priming selection ( • Ensure administration set is not soft key) is available only after the syringe and infusion type connected to patient.
Page 76
After priming is completed, reinstall the pressure sensing disc. OPTIONS Press key. SILENCE SYSTEM OPTIONS ENTER CLEAR CANCEL 2-12 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 77
Press CONFIRM and continue priming. >Press CONFIRM CONFIRM Invert pressure sensing disc so that patient side is up. Hold pressure sensing disc between fingers. Patient Side 2-13 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 78
Prime Set with Syringe CAUTION: Do not prime with disposable set connected to patient! Prime Volume 2 mL >Press and Hold PRIME EXIT PRIME 2-14 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 79
A A A disc is reinstalled, a prompt to reinstall pressure Re-install Pressure Sensing sensing disc displays. Disc. >Press CONFIRM CONFIRM 2-15 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 80
To deliver PRIME additional amounts, press the soft key again. Volume used during priming is displayed but not added to VTBI 2-16 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
After priming is completed, reinstall the pressure sensing disc. 2-17 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 82
Load syringe (see "Loading" procedure). If a pressure sensing disc is being used, do not install disc until priming is complete. Select syringe and infusion type (see "Programming" chapter). 2-18 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 83
PRIME soft key. Prime until fluid drips from end of tubing. Complete procedure (installing pressure sensing disc, if applicable, and exiting options menu). 2-19 Getting Started ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 84
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 2-20 Getting Started Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Confirm current profile or select a new profile. Enter patient identifier, if required. Prepare and load syringe/administration set (see "Getting Started"). Prime (see "Getting Started"). 2-21 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 86
If a default syringe list has been enabled and correct BD 30 mL ALL SYRINGES syringe cannot be found, press soft Monoject 30 mL key. >Select Syringe Size CONFIRM SYRINGES 2-22 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
• drug selected • weight entry (if required) • rate or dose entry • entry (Syringe Module: if other than All) VTBI 2-23 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 88
Once drug or IV fluid has been >Select Therapy Type confirmed, therapy indication appears in title bar. DRUG EXIT LIBRARY -- Continued on Next Page -- 2-24 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 89
DOSING mg/kg/h To change selection, press No soft key. UNITS BOLUSABLE BOLUS mg/kg DOSING UNITS >Press Yes or No -- Continued on Next Page -- 2-25 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 90
TIME hour UNITS DOSING mg/kg/h UNITS [Conc]: 1 mg/mL >Enter Patient Weight DRUG LIBRARY NEXT Verify parameters are correct and press soft key to confirm. 2-26 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 91
VTBI ♦ When is disabled and soft key is [Conc]: 1 mg/mL pressed, estimated available volume in syringe >Select VTBI displays. PAUSE SETUP BOLUS START 2-27 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 92
Message Display also shows either ↑↑↑ " " for a low dose or " " for a high dose. -- Continued on Next Page -- 2-28 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 93
2-29 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 94
If programmed continuous dose is weight-based, CONT- SETUP INUOUS programmed weight displays. If bolus dose is not weight-based, Not Used displays • PATIENT WEIGHT field. 2-30 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 95
Accept weight change? >Select Yes or No BOLUS e. Press soft key. To enter bolus dose, use numeric data entry keys. DURATION Press soft key. 2-31 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 96
If a Soft Limit is overridden, G icon displays. When G Midtown Hospital • Adult ICU soft key is pressed, all applicable out-of-range limits VTBI = 250 mL are listed. VOLUME AUDIO INFUSED ADJUST 2-32 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 97
START BOLUS To stop bolus and start continuous infusion, press Yes soft Alteplase Stroke key. Stop Bolus and Start Continuous infusion? >Press Yes or No 2-33 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 98
Bolus dose completed - module turned off: CHANNEL SELECT Press key. RESTORE Press soft key. NEXT Verify parameters and press soft key. BOLUS Verify infusion parameters and press soft key. 2-34 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 99
• rate or duration dose entry • entry VTBI Syringe Module: The option is disabled when an intermittent infusion is programmed. Press Guardrails Drugs soft key. 2-35 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 100
If selected drug is not weight-based, Not Used • PATIENT WEIGHT displays in field. -- Continued on Next Page -- 2-36 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 101
If a Soft Limit is overridden, G icon displays. When G Midtown Hospital • Adult ICU soft key is pressed, all applicable out-of-range limits VTBI = 250 mL are listed. VOLUME AUDIO INFUSED ADJUST 2-37 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 102
If a Soft Limit is overridden, G icon displays. When G • soft key is pressed, all applicable out-of-range limits are listed. 2-38 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 103
Press Guardrails IV Fluids soft key. Guardrails IV Fluids Oncology Packed Red Cells Press soft key next to IV Fluid to be delivered. Platelets >Select IV Fluid PAGE PAGE DOWN 2-39 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 104
Guardrails Fluid Setup entry keys. PRIMARY INFUSION RATE mL/h 1000 VTBI >Enter VTBI VOLUME PAUSE START DURATION -- Continued on Next Page -- 2-40 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 105
VTBI = 250 mL are listed. VOLUME AUDIO INFUSED ADJUST Syringe Module: Attach administration set to patient. RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART 2-41 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 106
DURATION VTBI ALL VTBI ♦ When is disabled and soft key is pressed, estimated available volume in syringe >Enter VTBI displays. RATE START PAUSE VOLUME 2-42 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 107
-- Continued on Next Page -- 2-43 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 108
Multiple doses from a single container are not possible. -- Continued on Next Page -- 2-44 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 109
Secondary Container Secondary Container Hanger Mini Bag Bottle Primary at least 9½" Container Primary at least 9½" Container 2-45 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 110
-based option for delivery of this infusion may >Select Drug appear. DRUG PAGE EXIT DOWN CALC • If applicable, multiple concentration listings for delivery of this infusion may appear. 2-46 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 111
Hard Limit for that care area, an audio alert sounds and a visual prompt appears before programming can continue. Infusion needs to be reprogrammed. -- Continued on Next Page -- 2-47 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 112
DURATION hh:mm 1200 DOSE = mg/m [Conc]: 136.8 mg/mL >Verify Secondary Clamp Open, Then Press START RATE PAUSE SETUP START VOLUME 2-48 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Prepare and load syringe/administration set (see "Getting Started"). Prime (see "Getting Started"). Start applicable infusion, as described in following procedures: Basic Infusion Continuous Infusion - Drug Calculation Bolus Dose 2-50 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
PAUSE DELAY OPTIONS Rate / Volume and Volume / Duration Infusions See "Primary Infusion - With Guardrails ® Suite MX Protection", "IV Fluid Infusion" procedure. 2-51 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 116
Press soft key for appropriate unit of measure for drug amount. To enter diluent volume, use numeric data entry keys. PATIENT WEIGHT Press soft key. 2-52 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 117
15. Verify parameters are correct and press soft key. Syringe Module: Attach administration set to patient. RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART -- Continued on Next Page -- 2-53 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 118
To enter bolus duration, use numeric data entry keys. TOTAL DOSE INFUSE AT • alternates with rate. START Verify parameters are correct and press soft key. 2-54 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
• drug to be infused is not listed in Drug Library, • primary infusion is running, and • a check valve administration set is being used. 2-55 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Secondary setup screen displays. START To resume secondary infusion, press soft key. Stopping Secondary and Returning to Primary ® See "Secondary Infusion - With Guardrails Suite MX Protection". 2-56 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
To enter desired parameter, use up/down arrows for rate titration, or numeric data entry keys. START Verify correct infusion parameter entry and press soft key. 2-57 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
401.1 08:00 2002-03-10 42.5 07:30 2002-03-11 478.1 08:00 2002-03-10 789.1 12:00 2002-03-10 >Select Channels to Clear or Press CLEAR ALL CLEAR MAIN PRI/SEC VOLUME SCREEN 2-58 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 123
If no key is pressed, main screen appears after seconds. The illustrated example is a Syringe Module display. A Pump SEC VOLUME Module display has a soft key. 2-59 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
>Select an Option or EXIT EXIT Channel Options 2 of 2 Channel Labels RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART >Select an Option or EXIT PAGE EXIT 2-60 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 125
To view additional labels, press a soft key next to a letter PAGE UP group to navigate through alphabet, and/or PAGE DOWN soft keys. 2-61 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
RATE (mL/h) RATE (mL/h) Anesthesia Mode CHANNEL SELECT CHANNEL SELECT PAUSE PAUSE CHANNEL CHANNEL RESTART RESTART See the PC Unit Section of this DFU. 2-62 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
CHANNEL RESTART RESTART ® Delay Options can be enabled at the time the Alaris System is configured for use. If Delay Options is enabled, a primary infusion can be programmed to be delayed for a specified period of time and a callback can be scheduled, as described in the following procedures.
Page 128
Delay period counts down on Main Display. If a Before callback has not been scheduled (see • "Scheduling a Callback" procedure), infusion automatically initiates at end of delay period. 2-64 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 129
(see "Scheduling a Callback" procedure), infusion automatically initiates at end of delay period. NOTE: If the current time has been previously confirmed, the Time of Day screen is not displayed. 2-65 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
♦ Yellow Standby Status Indicator flashes. DELAY COMPLETE ♦ scrolls in Message Display and appears on Main Display. -- Continued on Next Page -- 2-66 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
• Single dose piggybacking RATE (mL/h) systems employing check CHANNEL SELECT PAUSE CHANNEL valve sets are not designed RESTART for use with multidose containers. 2-68 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Pressure Limit - 550 mmHg Prime Set with Syringe >Select an Option or EXIT PAGE EXIT DOWN Start applicable infusion, as described in following procedures: Volume/Duration Enabled Volume/Duration Disabled 2-69 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 134
At completion of a multidose program, COMPLETE appears on Main Display. Syringe Module: Attach administration set to patient. RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART 2-70 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 135
To enter number of doses, press soft key and use numeric data entry keys. DELAY OPTIONS • If Delay Options is enabled, soft key appears. 2-71 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 136
2-72 Programming ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Press either Pump or Selectable pressure soft key. If Selectable is pressed, continue with next step; otherwise, Pressure Limit Selection proceed to last step. Pump Selectable >Select an Option or EXIT EXIT 2-73 Programming Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
CHANNEL SELECT Press key. after an infusion has started can result in a bolus to the patient. OPTIONS Press key. SILENCE SYSTEM OPTIONS ENTER CLEAR CANCEL 2-74 Programming Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 139
CONFIRM CONFIRM Verify correct pressure limit input and press soft key. Pressure Sensing Disc NOT Installed CHANNEL SELECT Press key. OPTIONS Press key. 2-75 Programming Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 140
• greater than mmHg – system adds to current pressure, to create a new alarm limit 2-76 Programming Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
To enter , press soft key and use numeric data VTBI entry keys. Open roller clamp. START To resume infusion, press soft key. 2-77 General Setup and Operation ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
® installed as part of a second Alaris System setup. Before changing the infusion line at the patient end, start the infusion and wait for fluid to drip from the end of the tubing.
• Discard if packaging is not intact or protector caps are unattached. 2-79 General Information ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 144
RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART CAUTION Before operating instrument, verify that administration set is free from kinks and installed correctly in instrument. 2-80 General Information ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
SELECT PAUSE CHANNEL RESTART ® • The Alaris System can be used for epidural administration of anesthetic and analgesic drugs. This application is only appropriate when using anesthetics and analgesics labeled for continuous epidural administration and catheters intended specifically for epidural use. Use only a Pump Module/Gemini Infusion System administration set or syringe set, without a ‘...
• Discard administration set per facility protocol. • For push medication (put instrument on hold), clamp tubing above port. • Flush port(s) per facility protocol. 2-82 General Information ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
• Do not leave slip luer syringes NOTE: unattended. Dry time is dependent on area temperature, humidity and ventilation. 2-83 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
In this closed position, flow is occluded. White Slide Clamp (shown in closed position) NOTE: "Safety clamp" is referred to on the device as "Flo-Stop". 2-84 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
When using a only the syringe size and type specified on the Main Display. syringe, Cardinal Health strongly The full list of permitted syringe models is dependent on the recommends using an extension set Syringe Module’s software version.
Alert. Drug label in Message ↑↑↑ Display provides an indicator (" " or " ") that dose is beyond current Soft Limit. 2-86 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 151
RESTART Drug Calculation Allows: ® • entry of drug dose for a continuous infusion ( System Alaris calculates correct flow rate to achieve desired dose), ® • entry of flow rate for a continuous infusion ( ystem Alaris calculates corresponding drug dose).
Page 152
Hospital-established Limits around total dose of infusion. Volume/Duration Allows a volume-to-be-infused ( ) and duration (infusion time) VTBI to be programmed. Flow rate is automatically calculated. 2-88 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 153
PAUSE CHANNEL RESTART Auto-Restart ® Part of Alaris System’s Downstream Occlusion Detection system designed to minimize nuisance, patient-side occlusion alarms. Allows system to automatically continue an infusion following detection of a patient-side occlusion if downstream pressure falls to an acceptable level within a -second "Checking Line"...
Page 154
• Auto-Restart: (See "Auto-Restart" definition.) ® In addition, Alaris System provides fluid-side occlusion detection. Secondary Infusions Dual rate sequential piggyback (secondary) infusions may be infused, with limits, at delivery rates and volumes independent of primary infusion parameters.
Page 155
Customizable Pressure Alarm Settings (see "Occlusion Pressure") Fast Start Pressure Tracking Pressure Tracking Dynamic current pressure display is only available when pressure sensing disc is inserted. 2-91 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 156
Instrument gives an alert and stops when an empty syringe is detected. Syringe Volume Detection System automatically detects fluid volume in a syringe when it is inserted. 2-92 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
When pressed, resumes operation of a previously paused or alarmed infusion on that module. Module Release Latch: Door Handle When pressed, allows module to be removed. 2-93 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 158
Lower Pumping Finger Lower Pressure Sensor Safety Clamp Device Recess Tubing Keeper Air-in-Line Sensor NOTE: "Safety clamp" is referred to on the device as "Flo-Stop". 2-94 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 159
When pressed, resumes operation of a previously paused or Module Release Latch: When pressed, alarmed infusion on that module. allows module to be removed. 2-95 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
The Pump Module’s range is Patient-side occlusion pressure mmHg and the Syringe Module’s set point is indicated by tick mark. range is mmHg. 1000 VOLUME AUDIO INFUSED ADJUST 2-96 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
(normalized for patient weight if specified by entering a patient weight). • Duration = / rate. VTBI • Drug concentration = drug amount / diluent volume. • Rate = / duration. VTBI 2-97 General Information ® Alaris System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
If the configuration settings need to be changed from the Factory default settings, reference the applicable Technical Service Manual or contact Cardinal Health Technical Support, for technical, troubleshooting, and preventive maintenance information. With the Profiles feature enabled, the settings are configured independently for each profile.
Occlusion Pressure Set Point: • With Disc mmHg mmHg in mmHg increments 1000 1000 Low, Medium, High • No Disc High Enabled - Disabled Priming Disabled 2-100 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
>85%) Temperature Range: 104° F 140° F 40° C 60° C Equipment Orientation: ® To ensure proper operation, Alaris System must remain in an upright position. Flow Rate Programming Increments (mL/h) Rate Range Increments: User Input Rates Device Calculated Rates (mL/h) 0.01...
Page 166
<30 Selectable Mode: User selected, mmHg in mmHg increments. Operating Principle: Positive displacement Rate Accuracy: ® Rate accuracy of Alaris System is at rates between mL/h and ±5% at rates mL/h, of the time with confidence, under conditions ±5.5% <1...
0.575 0.331 mmHg 1000 0.839 0.323 Alaris ® System has a back-off safety feature which, when enabled and a pressure sensing disc is in use, is designed to reduce bolus volume on occlusion release. -- Continued on Next Page --...
Page 168
Atmospheric Pressure: mmHg mmHg 4560 hPa) hPa) 6080 1013 Equipment Orientation: ® To ensure proper operation, Alaris System must remain in an upright position. 2-104 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 169
Without Pressure Sensing Disc: Three settings: Low, Medium, High With Pressure Sensing Disc: User selected, mmHg in mmHg increments. 1000 Operating Principle: Positive displacement 2-105 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 170
Syringe Type: BD 50 Temperature: ± ° Volume to be Infused Range (mL) Increments (mL) Programming Increments: 0.1 - 9.99 0.01 10 - 60 Weight: 2-106 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Authorized representative in European Community. EC REP RATE (mL/h) Pump Module CHANNEL SELECT PAUSE CHANNEL RESTART Drops per milliliter specification for product is identified on drop symbol. 2-107 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
They exhibit the delay in onset of delivery due to mechanical compliance and provide a visual representation of uniformity. Trumpet curves are derived from the second hour of this data. 2-108 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 173
In this case, the linear relationship of the plunger position and velocity to syringe volume and rate is verified, and is a function of the accuracy of the design. 2-109 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 174
+100 +300 pressures, the Syringe Module typically exhibits a long-term accuracy offset of approximately or less from the mean 0.2% value. 2-110 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Maximum rate error NOTE: The plot range has been increased to ±100% Overall rate error to allow visualization of the graph. Minimum rate error 2-111 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 176
Trumpet Curve at 999 mL/h (24 hrs) Trumpet Curve at 25 mL/h (72 hrs) Observation Interval (min) Observation Interval (min) Legend: Maximum rate error Overall rate error Minimum rate error 2-112 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 177
-1.8 -3.6 -1.9 -2.2 -4.1 -2.8 -5.3 -4.2 Observation Interval (min) Observation Interval (min) Legend: Maximum rate error Overall rate error Minimum rate error 2-113 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 178
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 2-114 General Information Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Troubleshooting and Maintenance General The Pump Module and Syringe Module Technical Service Manuals are available from Cardinal Health. They include routine service schedules, interconnect diagrams, component parts lists and descriptions, test procedures, and other technical information, to assist qualified service personnel in repair and maintenance of the instrument’s repairable...
"clicked" into place at Module Release Latch. If alarm is still present, replace module with an operational instrument. 2-116 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 181
Partial occlusion of patient side of Clear occlusion. Press CHANNEL SELECT IV line detected by Auto-Restart key, or press feature. START key and then soft key. 2-117 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 182
Pressure Disc Removed Reinsert pressure sensing disc and Pressure sensing disc removed. RESTART press key. Infusion stops on affected module. 2-118 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 183
CONTINUOUS INFUSION Syringe Installation of the problem. lever plunger flange clamp >Press CANCEL or Adjust Syringe CONFIRM -- Continued on Next Page -- 2-119 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 184
Drive Not Engaged Open and close plunger grippers Drive system disengaged during and syringe barrel clamp. Ensure operation. syringe is properly installed. 2-120 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Syringe Driver Head Error CONFIRM grippers open, senses excessive normal operation, press pressure being applied downward soft key. OCCLUSION on Drive Head. scrolls in Message Display. 2-121 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
SELECT START key and then soft key. CONFIRM Start time for next dose has Start of next dose passed. Press soft key. passed. 2-122 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Page 187
CONFIRM press ; or use a syringe type and size that system can automatically and correctly identify. 2-123 Troubleshooting and Maintenance Alaris ® System (with v9 Model 8000) DFU Pump and Syringe Modules Section...
Numeric Syringe Empty Syringe Empty Yes / Yes Inspection Requirements ® To ensure the Alaris System remains in good operating WARNING condition, both regular and preventive maintenance inspections Failure to perform these are required. Reference the System Maintenance software for inspections may result in improper detailed instructions.
Page 189
® Alaris PCA Module Model 8120 RATE (mL/h) CHANNEL SELECT PAUSE CHANNEL RESTART ® Alaris System (with v9 Model 8000) DFU, Section...
Page 191
3-39 ....................................Epidural Administration 3-41 ..................................Dose Request Cord 3-41 ....................................® Guardrails Suite MX 3-42 ..................................... ADMINISTRATION SET INFORMATION 3-42 ..............................COMPATIBLE SYRINGES 3-43 ..................................... ® Alaris System (with v9 Model 8000) DFU Table of Contents PCA Module Section...
Page 192
TROUBLESHOOTING AND MAINTENANCE GENERAL 3-55 ..........................................ALARMS, ERRORS, MESSAGES 3-55 ................................Definitions 3-56 ........................................Alarms 3-56 ........................................... Errors 3-58 ..........................................Messages 3-58 ........................................INSPECTION REQUIREMENTS 3-60 ................................. ® Alaris System (with v9 Model 8000) DFU Table of Contents PCA Module Section...
PC Unit Section of this ® • System Maintenance software (and its instructions) for Alaris System check-in, maintenance, and wireless configuration The PCA Module is intended for facilities that utilize syringe pumps for the delivery of medications or fluids. The PCA...
Dose Request Cord Attachment To detach Dose Request Cord: • Hold body of latching connector and pull straight away, without twisting or turning, from Dose Request Cord attachment. ® Alaris System (with v9 Model 8000) DFU Getting Started PCA Module Section...
For a list of compatible syringes, see "General Information", "Compatible Syringes". For a list of compatible administration sets, reference the Set Compatibility Card (provided separately). ® Alaris System (with v9 Model 8000) DFU Getting Started PCA Module Section...
Pull syringe barrel clamp out and hold. Rotate clamp to left (clockwise or counter clockwise) until it clears syringe chamber. Gently release clamp. RATE (mL/h) CHANNEL SELECT Syringe Barrel Clamp Open ® Alaris System (with v9 Model 8000) DFU Getting Started PCA Module Section...
Page 197
Rotate clamp to right (clockwise or counter clockwise) RATE (mL/h) until it lines up with syringe. CHANNEL Gently release clamp against syringe. SELECT PAUSE Syringe Barrel Clamp Closed ® Alaris System (with v9 Model 8000) DFU Getting Started PCA Module Section...
PROGRAM unlocking security door or interrupting current infusion. • locks security door. Key must be in this position to LOCK start an infusion. ® Alaris System (with v9 Model 8000) DFU Getting Started PCA Module Section...
(see "Service Information" in "Appendix" Section of this DFU). CHANNEL SELECT PROGRAM Press key. Key must be in RATE (mL/h) position. CHANNEL SELECT PAUSE CHANNEL RESTART PCA Module ® Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
If the syringe is not recognized, Syringe not recognized displays. WARNING Priming When priming: The Priming option can be enabled at the time the Alaris ® System is configured for use. The Priming selection ( PRIME • Ensure patient is not connected.
Page 201
• Volume used during priming is displayed but not added disposable set VTBI connected to patient! Prime Volume >P ress and Hold PRIME PRIME EXIT ® Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Unlock security door or set key to position. Confirm time of day or change time if necessary. Time of Day Current time: Change 08:00 Time >CONFIRM Time-of-Day EXIT CONFIRM ® 3-10 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Page 203
Advisory for that drug, a message appears. To CONFIRM continue programming, press soft key. NEXT 10. Verify parameters are correct and press soft key to confirm. 11. Prime syringe using Prime feature, if desired. ® Alaris System (with v9 Model 8000) DFU 3-11 Programming PCA Module Section...
When the PC Unit is in the Infusion Mode Selection, Infusion Setup or Bolus Setup screens, a patient dose request from the Dose Request Cord is handled as an unmet demand. ® 3-12 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Page 206
Time (in hours) associated with Max Limit is automatically entered based on setup in system configuration. Loading dose is included in volume infused but is not included in Max Limit. ® 3-14 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Yes soft key and use numeric data entry keys. CONT mg/h DOSE > > > > LIMIT LOADING None DOSE [Conc]: 1 mg/mL >Select Max Limit Option CHANGE MODE ® Alaris System (with v9 Model 8000) DFU 3-15 Programming PCA Module Section...
Page 208
Time (in hours) associated with Max Limit is automatically entered based on setup in system configuration. Loading dose is included in volume infused but is not included in Max Limit. ® 3-16 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Yes soft key and use numeric data entry keys. Morphine DOSE LOCKOUT minutes INTERVAL CONT mg/h DOSE > > > > LIMIT LOADING None DOSE [Conc]: 1 mg/mL >Select Max Limit Option CHANGE MODE ® Alaris System (with v9 Model 8000) DFU 3-17 Programming PCA Module Section...
Page 210
Time (in hours) associated with Max Limit is automatically entered based on setup in system configuration. Loading dose is included in but is not included in VTBI Max Limit. ® 3-18 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Dose: Main Display. [Conc]: 1 mg/mL When loading dose is complete, The Loading Dose >Press START • has Completed appears on Main Display. PRO- PAUSE START GRAM ® Alaris System (with v9 Model 8000) DFU 3-19 Programming PCA Module Section...
START GRAM DOSE PROGRAM Set key to position or enter -digit authorization CONFIRM code and press soft key. To enter dose value, use numeric data entry keys. ® 3-20 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Page 213
Programmed infusion resumes. GRAM DOSE NOTE: BOLUS DOSE soft key is only available once an infusion has begun in dose only, continuous infusion, or continuous infusion modes. ® Alaris System (with v9 Model 8000) DFU 3-21 Programming PCA Module Section...
CONFIRM MODE Select desired infusion mode. Continue programming. See applicable procedure: Setting Up PCA Dose Only Setting Up Continuous Infusion Only Setting Up PCA + Continuous Infusion ® 3-22 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Time Scale CLEAR ZOOM EXIT DETAIL HISTORY To continue and clear patient history, press Yes soft key. To cancel and return to patient history, press No soft key. ® 3-24 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Press Drug Event History soft key. Channel Options 1 of 2 PCA Pause Limits Infusion Modes Patient History Dose Request Setup Drug Event History >Select an Option or EXIT PAGE EXIT DOWN ® Alaris System (with v9 Model 8000) DFU 3-25 Programming PCA Module Section...
Channel Options 1 of 2 PCA Pause Limits Infusion Modes Patient History Dose Request Setup History Drug Event >Select an Option or EXIT PAGE EXIT DOWN ® 3-26 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Load syringe and administration set (see "Getting Started", "Preparing and Loading Syringe and Administration Set", "Loading Syringe and Administration Set"). Select syringe type and size (see "Preparing Infusion", "Selecting Syringe Type and Size"). ® Alaris System (with v9 Model 8000) DFU 3-29 Programming PCA Module Section...
Page 222
To change a restored parameter: a. Press applicable soft key. b. Enter desired parameter using numeric data entry keys. CONFIRM c. Press soft key. ® 3-30 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Option to change pressure limit can be selected: • after drug is selected, and before infusion mode is selected and infusion starts, or • after infusion starts. ® Alaris System (with v9 Model 8000) DFU 3-31 Programming PCA Module Section...
PCA Therapy. All programming, data entry and validation of PCA Pause Protocol parameters are performed by a healthcare professional according to hospital-defined protocol/procedure or a physician’s order. ® 3-32 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Start applicable infusion, as described in following procedures: Setting Up PCA Dose Only Setting Up Continuous Infusion Only Setting Up PCA Dose + Continuous Infusion Setting Loading Dose Only ® Alaris System (with v9 Model 8000) DFU 3-33 Programming PCA Module Section...
Press PCA Pause Limits soft key. Channel Options 1 of 2 PCA Pause Limits Infusion Modes Patient History Dose Request Setup History Drug Event >Select an Option or EXIT PAGE EXIT DOWN ® 3-34 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
Page 227
® Suite PCA PAUSE LIMITS SPO2 ETCO2 must be lower than the ALARM LIMITS PCA PAUSE LIMITS . A prompt is provided if the must be modified. ® Alaris System (with v9 Model 8000) DFU 3-35 Programming PCA Module Section...
NOTES: Disabling from this screen discontinues the PCA Pause feature only, without interrupting monitoring functionality. Once disabled, the alarm limits are grayed out and are not editable. ® 3-36 Alaris System (with v9 Model 8000) DFU Programming PCA Module Section...
General Setup and Operation System Start-Up / Setup See the PC Unit Section of this DFU "General Setup and Operation", for various system start-up and setup procedures. ® Alaris System (with v9 Model 8000) DFU 3-37 General Setup and Operation PCA Module Section...
Page 230
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® 3-38 Alaris System (with v9 Model 8000) DFU General Setup and Operation PCA Module Section...
"gravity only" systems, affecting their performance. Hospital/facility personnel must ensure the performance of the common site is satisfactory under these circumstances. • Each time the Alaris System is turned on, verify and/or ® set the monitoring mode, resistance alert, and/or pressure alarm limit.
Page 232
• Discard if packaging is not intact or protector caps are unattached. CAUTION Before operating instrument, verify that administration set is free from kinks and installed correctly in instrument. ® 3-40 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
PCA Module used for epidural drug delivery be clearly differentiated from those used for other types of administration. ® • The Alaris System can be used for epidural administration of anesthetic and analgesic drugs. This application is only appropriate when using anesthetics and analgesics labeled for continuous epidural administration and catheters intended specifically for epidural use.
Do not resterilize. • Discard administration set per facility protocol. • push medication (put instrument on hold), clamp tubing above port. • Flush port(s) per facility protocol. ® 3-42 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
Allows an additional amount of medication to be programmed once infusion has begun. Current infusion resumes following delivery of a bolus dose. Continuous Dose Basal rate dose. ® Alaris System (with v9 Model 8000) DFU 3-43 General Information PCA Module Section...
Page 236
Once drugs are in Profile Drug Library, Max Accumulated Dose Limit cannot be changed. • Applies to all drug setups within Profile PCA Drug Library. ® 3-44 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
Page 237
Enables a patient to self-administer a bolus infusion to be delivered at programmed lockout intervals through Dose Request Cord. When programmed in +continuous mode, continuous infusion resumes following dose. ® Alaris System (with v9 Model 8000) DFU 3-45 General Information PCA Module Section...
Page 238
An optional hospital-defined therapy or clinical indication for delivery of that infusion. Different Limits can be defined for same medication with different therapeutic indications. Time Window (h) hours. ® 3-46 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
Repeat for other operating modules to power off each module. Dose Request Cord Attachment Dose Request Cord ® Alaris System (with v9 Model 8000) DFU 3-47 General Information PCA Module Section...
See the PC Unit Section of this DFU for system configurable settings. The configuration settings are selected during Data Set ® development and then uploaded to the Alaris System as part of the data. With the Profiles feature enabled, the settings are configured independently for each profile.
Page 241
These values are configured in the Module settings within the Editor Software and can be changed by the clinician by accessing Channel Options on the PCA Module. ® Alaris System (with v9 Model 8000) DFU 3-49 General Information PCA Module Section...
>85%) mmHg mmHg Atmospheric Pressure: 4560 hPa) hPa) 6080 1013 Equipment Orientation: To ensure proper operation, Alaris ® System must remain in an upright position. ® 3-50 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
Page 243
"Trumpet and Start-Up Curves" for data on how these factors influence rate accuracy. Shock Protection: Type , Defibrillator Proof (PCA Module) Type , Defibrillator Proof (Dose Request Cord) ® Alaris System (with v9 Model 8000) DFU 3-51 General Information PCA Module Section...
DEHP DEHP Product is latex-free. LATEX Approximate administration set priming volume. » XX ml Expiration date for product is identified near hour glass symbol. ® 3-52 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
As the observation window is reduced, short-term fluctuations have greater effect, as represented ® Alaris System (with v9 Model 8000) DFU 3-53 General Information PCA Module Section...
Page 246
Trumpet and Start-Up Curves (Continued) Legend: Maximum rate error Overall rate error Minimum rate error ® 3-54 Alaris System (with v9 Model 8000) DFU General Information PCA Module Section...
Troubleshooting and Maintenance General The PCA Module Technical Service Manual is available from Cardinal Health. It includes routine service schedules, interconnect diagrams, component parts lists and descriptions, test procedures, and other technical information, to assist qualified service personnel in repair and maintenance of the instrument’s repairable components.
PCA Pause Alarm CONFIRM decline in respiratory status. policy. Press once patient status and monitoring values have been addressed. RESTART Press key per hospital policy. ® 3-56 Alaris System (with v9 Model 8000) DFU Troubleshooting and Maintenance PCA Module Section...
Page 249
Securely lock plunger grippers over syringe plunger. Drive system disengaged during Open and close plunger grippers Drive Not Engaged operation. and syringe barrel clamp. Ensure syringe is properly installed. ® Alaris System (with v9 Model 8000) DFU 3-57 Troubleshooting and Maintenance PCA Module Section...
, press specified. Infusion paused until PROGRAM soft key, and unlock time limit has expired. door or enter Authorization Code applicable for current Security Access Level. ® 3-58 Alaris System (with v9 Model 8000) DFU Troubleshooting and Maintenance PCA Module Section...
Page 251
CONFIRM press , or use a syringe type and size that system can automatically and correctly identify. ® Alaris System (with v9 Model 8000) DFU 3-59 Troubleshooting and Maintenance PCA Module Section...
Inspection Requirements ® To ensure the Alaris System remains in good operating WARNING condition, both regular and preventive maintenance inspections Failure to perform these are required. Reference the System Maintenance software for inspections may result in improper detailed instructions. instrument operation.
Module sensor and cable Compatibility Cards • Module Technical Service Manual ® • System Maintenance software (and its instructions) for Alaris System check-in, maintenance, and wireless configuration Modules are indicated for continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin ( ) and pulse rate measured by an sensor.
Patient Cable sensor’s directions for use for detailed instructions. Attach patient Attach sensor cable here. here. Attach sensor to patient. Reference sensor’s directions for use for detailed instructions. ® Alaris System (with v9 Model 8000) DFU Getting Started Modules Section...
If sensor is not attached during message display, module goes into sleep mode. To begin monitoring once module is in this mode, press key. MONITOR Ensure sensor's red is on. ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
Setting Alarm Limits Press soft key. Adult LIMITS SPO2 Main Mode % SPO2 PULSE RATE SatSeconds (% to Alarm) >Press ENTER to Retain Current View MAIN TREND LIMITS SCREEN ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
PULSE RATE SatSeconds (% to Alarm) >Press ENTER to Retain Current View MAIN TREND LIMITS SCREEN To navigate from page to page, press PAGE UP PAGE soft keys. DOWN ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
Page 262
To view the latest data, return to the Trend Data view. displays if an alarm limit is reached. If no rate values are available for the time SPO2 PULSE period displayed, dashes (---) are displayed. ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
To view the latest data, return to the Trend Data view. displays if an alarm limit is reached. If no rate values are available for the time SPO2 PULSE period displayed, dashes (---) are displayed. ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
(profile option not being used for programming). Press Limit Mode soft key. Channel Options Limit Mode: Adult Pulse Beep Volume: Off SatSeconds Limits: >Select an Option or EXIT EXIT ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
The pulse beep must be on to test the volume level. To turn pulse beep on, press the Louder soft key and adjust as needed (steps 2b and 2c). ® Alaris System (with v9 Model 8000) DFU Programming Modules Section...
• Maximum: Improved low perfusion performance. Maximum >Select an Option or EXIT EXIT NOTE: The sensitivity mode displays on the display only SPO2 Main when Maximum is selected. ® Alaris System (with v9 Model 8000) DFU 4-11 Programming Modules Section...
Page 268
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® 4-12 Alaris System (with v9 Model 8000) DFU Programming Modules Section...
General Setup and Operation System Start-Up / Setup See the PC Unit Section of this DFU, "General Setup and Operation", for various system start-up and setup procedures. ® Alaris System (with v9 Model 8000) DFU General Setup and Operation 4-13 Modules Section...
Page 270
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® 4-14 Alaris System (with v9 Model 8000) DFU General Setup and Operation Modules Section...
Module in any position that may cause it to fall onto the patient. • Respond immediately to system alarms; patient monitoring may cease under certain alarm conditions. ® Alaris System (with v9 Model 8000) DFU 4-15 General Information Modules Section...
• Before use, read the sensor directions for use, including all warnings, cautions and instructions. CAUTION Do not immerse or dampen the sensor or cable. Clean per manufacturer’s instructions. ® 4-16 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
For a list of compatible sensors and cables, reference the Sensor and Cable Compatibility Card (provided separately). ® Alaris System (with v9 Model 8000) DFU 4-17 General Information Modules Section...
Tabular display of and pulse rate. Display shows average high and low values, and alarm conditions for time period displayed. Up to hours of data is stored. ® 4-18 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
Page 275
A SatSeconds limit can be set to allow monitoring of below selected low alarm limit for a period of time before an audible alarm sounds. ® Alaris System (with v9 Model 8000) DFU 4-19 General Information Modules Section...
Page 276
Signal feature is related to proper sensor application, adequate arterial signal and intensity of ™ motion. Use Signal feature to verify optimal sensor placement. ® 4-20 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
Repeat for other operating modules to power off each module. Module Release Latch: When pressed, allows module to be removed. Patient Cable Connector ® Alaris System (with v9 Model 8000) DFU 4-21 General Information Modules Section...
SCREEN Adult SPO2 Main Mode % SPO2 (8220) PULSE RATE Pleth Waveform PI = 10.56 Signal I.Q.™ >Press ENTER to Retain Current View MAIN TREND LIMITS SCREEN ® 4-22 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
If the configuration settings need to be changed from the Factory default settings, reference the applicable Technical Service Manual or contact Cardinal Health Technical Support, for technical, troubleshooting, and preventive maintenance information. With the Profiles feature enabled, the settings are configured independently for each profile.
Atmospheric Pressure: mmHg mmHg 4560 hPa) hPa) 6080 1013 Fluid Ingress Protection: , Drip Proof IPX1 Mode of Operation: Continuous Shock Protection: Type Weight: lbs ( 0.91 ® 4-24 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
Page 281
100% 1% S Adults, Pediatrics: digits Adults, Pediatrics: ±3 digits; Adults, Pediatrics, digits; ±2 ±2 Neonates: digits Neonates Neonates: digits ±3 ±3 Display Update Period: Approximately second. ® Alaris System (with v9 Model 8000) DFU 4-25 General Information Modules Section...
Page 282
100% S co–oximeter and monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses of the population. ® 4-26 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
• Nail aberrations, nail polish, fungus, etc. Remove nail polish and/or move sensor to an unaffected site. • Placement is too close to electrosurgery equipment. • Defibrillation. ® Alaris System (with v9 Model 8000) DFU 4-27 General Information Modules Section...
Page 284
There is an arterial occlusion proximal to sensor. • Placement is too close to electrosurgery equipment. NOTE: Exposure to excessive illumination can be corrected by covering the sensor with a dark or opaque material. ® 4-28 Alaris System (with v9 Model 8000) DFU General Information Modules Section...
Troubleshooting and Maintenance General Module Technical Service Manual is available from Cardinal Health. It includes routine service schedules, interconnect diagrams, component parts lists and descriptions, test procedures, and other technical information, to assist qualified service personnel in repair and maintenance of the instrument’s repairable components.
Page 286
Sensor not properly attached to Attach sensor to patient cable patient cable or patient cable not or attach patient cable to properly attached to Module. Module. ® 4-30 Alaris System (with v9 Model 8000) DFU Troubleshooting and Maintenance Modules Section...
Low signal quality being measured. Check sensor. Confirm correct sensor placement. Move sensor to a better perfused site. Inspection Requirements To ensure the Alaris ® System remains in good operating WARNING condition, both regular and preventive maintenance inspections Failure to perform these are required.
Page 289
® Alaris EtCO Module Model 8300 EtCO (mm Hg) RR (breaths/min) CHANNEL SELECT MONITOR CHANNEL ® Alaris System (with v9 Model 8000) DFU, Section...
Page 291
TROUBLESHOOTING AND MAINTENANCE GENERAL 5-27 ..........................................ALARMS AND MESSAGES 5-27 ..................................Definitions 5-27 ........................................Audio Characteristics 5-28 ..................................... Alarms 5-28 ........................................... Messages 5-29 ........................................INSPECTION REQUIREMENTS 5-30 ................................. ® Alaris System (with v9 Model 8000) DFU Table of Contents EtCO Module Section...
Page 292
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Table of Contents EtCO...
CAUTION • PC Unit Section of this DFU • System Maintenance software (and its instructions) for Alaris ® System check-in, maintenance, and wireless configuration Module is a capnograph indicated for continuous, noninvasive monitoring of end tidal carbon dioxide (...
Module performance, resulting in inaccurate readings. For a list of compatible disposables, reference ® the Microstream Disposable Compatibility Card (provided separately). ® Alaris System (with v9 Model 8000) DFU Getting Started EtCO Module Section...
Gas scavenging system Release door. NOTE: The gas outlet is located on the lower right corner of the instrument and is marked with a gas outlet symbol ( ® Alaris System (with v9 Model 8000) DFU Getting Started EtCO Module Section...
Page 296
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Getting Started EtCO Module Section...
Channel Message display until and respiratory rate readings stabilize (up to seconds). CHANNEL SELECT MONITOR CHANNEL To change settings, see "Setting Alarm Limits" procedure. -- Continued on Next Page -- ® Alaris System (with v9 Model 8000) DFU Programming EtCO Module Section...
HIGH breaths/min seconds BREATH FiCO2 High : 10 mmHg >Press CONFIRM to Apply Changes FICO2 CONFIRM LIMIT To return to Main Display, press soft key. MAIN SCREEN ® Alaris System (with v9 Model 8000) DFU Programming EtCO Module Section...
Move Cursor. PAGE PAGE ETCO2 ZOOM DOWN MAIN To return to display, press soft ETCO2 Main ETCO2 MAIN key. To return to Main Display, press soft key. MAIN SCREEN ® Alaris System (with v9 Model 8000) DFU Programming EtCO Module Section...
TIME 08:04 08:05 • Return to display. ETCO2 MAIN 08:06 ZOOM: minutes • Return to Main Display. >Press UP/DOWN Keys to Move Cursor. ETCO2 PAGE ZOOM MAIN DOWN ® Alaris System (with v9 Model 8000) DFU Programming EtCO Module Section...
>Select an Option or EXIT • mmHg: Displays a waveform for full value EXIT range, mmHg. 0 – 99 To return to display, press soft key. ETCO2 Main EXIT ® 5-10 Alaris System (with v9 Model 8000) DFU Programming EtCO Module Section...
Waveform Time Scale 5 sec 10 sec To leave Waveform Time Scale unchanged and return to display, press soft key. ETCO2 Main EXIT >Select an Option or EXIT EXIT ® Alaris System (with v9 Model 8000) DFU 5-11 Programming EtCO Module Section...
Page 304
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® 5-12 Alaris System (with v9 Model 8000) DFU Programming EtCO Module Section...
General Setup and Operation System Start-Up / Setup See the PC Unit Section of this DFU, "General Setup and Operation", for various system start-up and setup procedures. ® Alaris System (with v9 Model 8000) DFU 5-13 General Setup and Operation EtCO...
Page 306
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® 5-14 Alaris System (with v9 Model 8000) DFU General Setup and Operation EtCO Module Section...
Module performance, resulting in inaccurate readings. For a list of compatible disposables, reference ® the Microstream Disposable Compatibility Card (provided separately). ® Alaris System (with v9 Model 8000) DFU 5-15 General Information EtCO Module Section...
Microstream ® Disposables, contact Oridion at http://www.oridion.com or 1-888-ORIDION For a list of compatible disposables, reference the Sensor and Cable Compatibility Card (provided separately). ® 5-16 Alaris System (with v9 Model 8000) DFU General Information EtCO Module Section...
Tabular display of and respiratory rate. Display shows average, high and low values and alarm conditions for time period displayed. Up to hours of data is stored. ® Alaris System (with v9 Model 8000) DFU 5-17 General Information EtCO Module Section...
Microstream ® Disposable Connector ( Gas Inlet Gas Exhaust: Gas scavenging system connection. Module Release Latch: When pressed, allows module to be removed. ® 5-18 Alaris System (with v9 Model 8000) DFU General Information EtCO Module Section...
If the configuration settings need to be changed from the Factory default settings, reference the applicable Technical Service Manual or contact Cardinal Health Technical Support, for technical, troubleshooting, and preventive maintenance information. With the Profiles feature enabled, the settings are configured independently for each profile.
Measures and reports partial pressures of in the range of mmHg at sea level. values are calculated for all valid breaths. Dimensions: 3.3" W 8.9" H 5.5" D 22.6 ® 5-20 Alaris System (with v9 Model 8000) DFU General Information EtCO Module Section...
Page 313
Isoflurane Sevoflurane Internal Power Source: Operating time (fully charged): hours Measurement Range: mmHg mmHg Respiratory Rate: Mode of Operation: Continuous Shock Protection: Type , Defibrillator Proof ® Alaris System (with v9 Model 8000) DFU 5-21 General Information EtCO Module Section...
Examples of alternate means, then ensure these conditions are as follows: Module is functioning correctly. • Leaks or internal venting of sampled gas may affect accuracy. ® 5-22 Alaris System (with v9 Model 8000) DFU General Information EtCO Module Section...
The following graphs are examples of common problems identifiable through waveform analysis. These are examples only and do not represent all potential abnormal waveforms. Abnormal waveforms are not always associated with alarms. ® Alaris System (with v9 Model 8000) DFU 5-23 General Information...
Page 316
Numerical values continue to be displayed on both the Module and PC Unit. ® 5-24 Alaris System (with v9 Model 8000) DFU General Information EtCO Module Section...
In high humidity environments or extended monitoring periods ( hours), only Microstream ® Disposables designed for those instances should be used. ® Alaris System (with v9 Model 8000) DFU 5-25 General Information EtCO Module Section...
Page 318
(Pb - Where: is fractional concentration of in dry gas and % CO Pb is ambient pressure. is partial pressure of BTPS ® 5-26 Alaris System (with v9 Model 8000) DFU General Information EtCO Module Section...
Troubleshooting and Maintenance General Module Technical Service Manual is available from Cardinal Health. It includes routine service schedules, interconnect diagrams, component parts lists and descriptions, test procedures, and other technical information, to assist qualified service personnel in repair and maintenance of the instrument’s repairable components.
Patient Not Detected Monitor or Channel Select key Assess patient condition. Check pressed and patient not detected. disposable. ® Alaris System (with v9 Model 8000) DFU 5-29 Troubleshooting and Maintenance EtCO Module Section...
Inspection Requirements ® To ensure the Alaris System remains in good operating WARNING condition, both regular and preventive maintenance inspections Failure to perform these are required. Reference the System Maintenance software for inspections may result in improper detailed instructions. instrument operation.
Page 323
® Alaris Auto-ID Module Model 8600 R E A D Y S C A N CA NC EL ® Alaris System (with v9 Model 8000) DFU, Section...
Page 325
6-14 ................................Specifications 6-14 ......................................Symbology 6-16 ........................................Symbols 6-16 ........................................TROUBLESHOOTING AND MAINTENANCE GENERAL 6-17 ..........................................ALARMS, ERRORS, MESSAGES 6-17 ................................INSPECTION REQUIREMENTS 6-17 ................................. ® Alaris System (with v9 Model 8000) DFU Table of Contents Auto-ID Module Section...
Page 326
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Table of Contents Auto-ID Module Section...
• PC Unit Section of this ® • System Maintenance software (and its instructions) for Alaris System check-in, maintenance, and wireless configuration The addition of the Auto-ID Module to the Alaris ® System ® combines Guardrails Suite with dose limit technology and bar code technology to provide a new level of medication safety.
Page 328
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Getting Started Auto-ID Module Section...
Cardinal Health. Using other accessories may result in increased emissions or ® decreased immunity of the Alaris System. Power on PC Unit. To select New Patient?, press Yes soft key. To accept current profile, press Yes soft key.
• Always verify that information displayed on the PC Unit matches NOTE: scanned data. Patient may be entered manually using the PC Unit keypad (see PC Unit Section of this DFU). ® Alaris System (with v9 Model 8000) DFU Programming Auto-ID Module Section...
Auto-ID application. This feature is designed to ensure that only clinicians with a bar code on ® their badge can program the Alaris System. When this feature is enabled, the PC Unit automatically enables tamper resist mode upon power on and minutes after programming is completed.
CHANNEL SELECT Program infusion (see applicable module-specific Section of this DFU). NOTES: The Alaris ® System determines if the module selected is appropriate for the scanned medication type. If the selection is not appropriate (for example, a bag was scanned but a...
General Setup and Operation System Start-Up / Setup See the PC Unit Section of this DFU, "General Setup and Operation", for various system start-up and setup procedures. ® Alaris System (with v9 Model 8000) DFU General Setup and Operation Auto-ID Module Section...
Page 336
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® 6-10 Alaris System (with v9 Model 8000) DFU General Setup and Operation Auto-ID Module Section...
• Carefully locate the hand-held scanner to reduce the possibility of patient entanglement or strangulation. • Use only the hand-held external scanner supplied by Cardinal Health. Using other accessories may result in increased emissions or decreased immunity of the Alaris ®...
Hand-Held Scanner The hand-held external scanner supplied by Cardinal Health is WARNINGS the only hand-held scanner approved for use with the Auto-ID • Do not open the hand-held Module. scanner case. If the case is opened, an electrical shock hazard and possible exposure to...
If the configuration settings need to be changed from the Factory default settings, reference the applicable Technical Service Manual or contact Cardinal Health Technical Support, for technical, troubleshooting, and preventive maintenance information. With the Profiles feature enabled, the settings are configured independently for each profile.
Page 341
( 1 ±0.1 436.5 ±43.65 Hand-Held Scanner Dimensions: 3.25" W 7.25" H 4.25" L cm L) 18.4 10.8 Housing: flammability rating UL 94V0 Weight: oz ( ® Alaris System (with v9 Model 8000) DFU 6-15 General Information Auto-ID Module Section...
Trioptic Code Code Pharmaceutical ( Micro PARAF PDF417 PDF417 Code Symbols See the PC Unit Section of this DFU for system symbols. Input. Hand-held connection point. ® 6-16 Alaris System (with v9 Model 8000) DFU General Information Auto-ID Module Section...
General Alaris ® System Technical Service Manuals are available from Cardinal Health. They include routine service schedules, interconnect diagrams, component parts lists and descriptions, test procedures, and other technical information, to assist qualified service personnel in repair and maintenance. Maintenance procedures are intended to be performed only by qualified personnel, using the Service Manual and System Maintenance software.
Page 345
Nurse Call Accessory Model 8010 Max Load 30V/1A NURSE CALL ® Alaris System (with v9 Model 8000) DFU, Section...
Page 347
NURSE CALL JACK AND CABLE ................................Nurse Call Location ....................................Nurse Call Cable Connected ................................SPECIFICATIONS ......................................TROUBLESHOOTING AND MAINTENANCE GENERAL ..........................................ALARMS, ERRORS, MESSAGES ................................® Alaris System (with v9 Model 8000) DFU Table of Contents Nurse Call Accessory Section...
Page 348
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Table of Contents Nurse Call Accessory Section...
) instructions and information. It is used in 8010 Read all instructions, including conjunction with: those for the PC Unit and attached ® module(s), before using the Alaris • Module-specific Sections of this System. • PC Unit Section of this Communications Interface (CI) Board Accessory Section of this •...
Page 350
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Introduction Nurse Call Accessory Section...
• The nurse call feature should not be used as the primary source of alarm notification. The audible and visual ® alarms of the Alaris System, used in conjunction with clinical signs and symptoms, are the primary sources providing notification that an alarm condition exists.
Specifications Contact Closure Rating: 30 VDC maximum at 1A maximum Drip Proof Protection: To ensure Alaris ® System fluid ingress rating, Nurse Call jack is covered/ sealed when not in use. Isolation: Nurse Call contacts are isolated from chassis up to 500 VAC.
Page 354
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU General Information Nurse Call Accessory Section...
General Alaris ® System Technical Service Manuals are available from Cardinal Health. They include routine service schedules, interconnect diagrams, component parts lists and descriptions, test procedures, and other technical information, to assist qualified service personnel in repair and maintenance. Maintenance procedures are intended to be performed only by qualified personnel, using the Service Manual and System Maintenance software.
Page 357
Communications Interface (CI) Board Accessory Model 8012 ® Alaris System (with v9 Model 8000) DFU, Section...
Page 359
........................................GENERAL INFORMATION CI BOARD ACCESSORY AND CI BOARD LED ..........................WIRELESS NETWORK INDICATOR ................................ TROUBLESHOOTING AND MAINTENANCE GENERAL ..........................................CI BOARD AND ALARIS ® SERVER CONNECTIONS ........................ALARMS, ERRORS, MESSAGES ................................® Alaris System (with v9 Model 8000) DFU...
Page 360
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Table of Contents CI Board Accessory Section...
Alaris ® System, Alaris ® Server and Mobile Systems be established or is lost. The Alaris ® Manager. Server and CI Board are designed to minimize these incidents but cannot ® Combined use of the Alaris System, CI Board, Auto-ID eliminate them.
Page 362
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU Introduction CI Board Accessory Section...
Flashes green when Alaris CI Board Accessory System is powered up. WIRELESS NETWORK INDICATOR Wireless Network Indicator When illuminated, indicates Alaris ® ® System is connected to Alaris Server SILENCE or Mobile Systems Manager. When SYSTEM OPTIONS blinking, indicates data transfer. ENTER...
Page 364
T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K ® Alaris System (with v9 Model 8000) DFU General Information CI Board Accessory Section...
When an Alaris ® Server or Mobile Systems Manager connection is made, the Wireless Network Indicator on the PC Unit lights up. If connection to the Alaris ® Server is interrupted, the indicator light is extinguished. Some of the causes for a communications failure include: ®...
A soft-bristled brush may be used to clean hard to reach and narrow areas. For sanitizing or antibacterial treatment, use bleach solution and water. Use a soft cloth dampened with water to rinse off cleaning solution. Alaris ® System (with v9 Model 8000) DFU Maintenance Appendix Section...
Data Set is loaded. Technical Support Technical support, service information, applications, and manuals may be obtained by contacting a Cardinal Health representative. When submitting any request for service, include: • model number • a description of difficulty experienced •...
Furthermore, this warranty shall not apply to, and Cardinal Health shall not be ® responsible for, any loss or damage arising in connection with the purchase or use of any Alaris System product which has been: repaired by anyone other than an authorized Cardinal Health Service Representative;...
Page 372
Maintenance ® Alaris System (with v9 Model 8000) DFU Appendix Section...
The authority to operate this system is conditioned by the requirement that no modifications are made to the system unless the changes or modifications are expressly approved by Cardinal Health, Inc. This Class digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulation.
Page 374
The type radio card, if provided, is manufactured by 4137 Symbol Technologies, Inc., Holtsville, N.Y. 11742 Tables: The Alaris ® System is intended for use in the electromagnetic environments specified in the following tables. Table 1 Electromagnetic Emissions Emissions Test...
Page 375
Compliance (Continued) Electromagnetic Environment (Continued) ® Alaris System (Continued) Table 2 Electromagnetic Immunity Emissions Test IEC 60601-1-2 Compliance Electromagnetic Environment - Guidance Test Level Level IEC 61000-4-2 ±6 kV contact ±8 kV contact Floors should be wood, concrete, or ceramic tile.
Page 376
(>95% dip in U commercial or hospital environment. Voltage Dips, Short for 0.5 cycle for 0.5 cycle ® Interruptions, and If continued operation of Alaris System is Voltage Variations required during power mains interruptions, it is 40% U 40% U recommended that Alaris ®...
Page 377
3 Vrms 10 Vrms Portable and mobile RF communications equipment ® Conducted RF 150 kHz - 80 MHz should be used no closer to Alaris System (including cables) than recommended separation 10 V/m IEC 61000-4-3 3 V/m distance calculated from equation applicable to Radiated RF 80 MHz - 2.5 GHz...
Page 378
Reduce the potential for electromagnetic interference by maintaining a minimum distance between ® portable and mobile communications equipment (transmitters) and the Alaris System as recommended in this table, based on the maximum output power of the communications equipment. For transmitters rated at a maximum output power not listed in this table, the recommended separation...
Page 379
To assess the electromagnetic environment due to fixed transmitters, an electromagnetic ® site survey should be considered. If the measured field strength in the location in which the Alaris System ® is used exceeds the applicable...
Page 380
Cet appareil numérique de la Classe respecte toutes les exigences du Reglement sur le Matériel Brouilleur du Canada. This Class digital device meets the requirements of the International community. A-12 Alaris ® System (with v9 Model 8000) DFU Regulations and Standards Appendix Section...
Page 381
Compliance (Continued) Federal Aviation Regulations The following Alaris ® products have received a Statement of Compliance with Federal Aviation Regulations for use as a "Portable Electronic Device Aboard Aircraft". This is pursuant to the Advisory Circular No. and attested by an...
Trademarks ® ® ® Alaris , Guardrails and SmartSite are registered trademarks ® of Cardinal Health, Inc. or one of its subsidiaries. LNOP ® ® ® Masimo , Signal Extraction Technology and Signal ® are trademarks of Masimo Corporation. Nellcor I.Q.