Alarms - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Alarm
Battery Discharged
Channel Disconnected
Very Low Battery <5 minutes to
system shutdown
Error
Audio System Error
Channel Error
Alaris
System Directions for Use
®
PC Unit Section
Alarms, Errors, Messages (Continued)

Alarms

Operation of all modules stopped
due to insufficient battery charge.
Module(s) disconnected while in
operation or have a communication
problem.
Battery has
power at current power
consumption rate before operation
stops.
Errors
Main speaker failure.
Error detected. Operation stops on
affected module.
Meaning
minutes or less of
5
Meaning
Response
Connect AC power cord to power
source; alarm will be silenced.
Press
key on affected
RESTART
module to continue operation of
paused modules.
To silence alarm and clear
message from screen, press
soft key. Reattach
CONFIRM
module, if desired, ensuring it is
securely "clicked" into place at
Module Release Latch. If alarm is
still present, replace module with
an operational instrument.
Connect
power cord to power
AC
source; alarm will be silenced.
Response
1. Visually check alarm status to
determine whether or not an
operational alarm also needs to
be addressed (red Alarm Status
Indicator would be lit).
2. Press
SYSTEM ON
down system. Replace PC Unit
with an operational instrument.
Service by qualified personnel
is required.
To silence alarm and continue
operation of unaffected modules,
press
soft key . Replace
CONFIRM
module with an operational
instrument, as required. Service by
qualified personnel is required.
Troubleshooting and Maintenance
key to power
1-37