Syringe Module (Continued)
Alarm
Syringe Empty
Syringe Adjustment Alarms
When a syringe installation problem is detected, a visual
signal is displayed. Text in the display blinks to indicate the
location of the problem.
-- Continued on Next Page --
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Alarms, Errors, Messages (Continued)
Alarms (Continued)
Syringe is empty.
If syringe is not empty, other
possibilities are:
• Pressure sensing disc
inappropriate/defective.
• Syringe plunger travel
impeded.
• Pressure transducer defective.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Meaning
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Troubleshooting and Maintenance
Response
Set up new infusion or press
key.
CHANNEL OFF
Verify appropriate pressure sensing
disc is in use and functioning
properly.
Verify syringe plunger movement is
unimpeded.
If syringe is not empty and above
actions do not correct alarm,
contact qualified service
personnel.
A A A
Drug Calculation
A
Heparin
CONTINUOUS INFUSION
Syringe Installation
lever
plunger
flange
clamp
>Press
CANCEL
or Adjust
Syringe
CONFIRM
2-113