Stopping A Loading, Pca Or Bolus Dose; Changing Programming Parameters During An Infusion - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Stopping a Loading, PCA or Bolus Dose

1.
Press
CHANNEL SELECT
2.
Press
STOP LOAD, STOP PCA
applicable.
3.
To stop dose and resume current program, press Yes soft
key.
NOTE:
Available soft key and stop confirmation screen are
dependent on the type of dose currently infusing and current
infusion mode.

Changing Programming Parameters During an Infusion

1.
Press
CHANNEL SELECT
2.
Press
PROGRAM
3.
Set key to program position or if Authorization Code is
enabled, enter
4
4.
Press
CHANGE MODE
3-22
Programming
Infusion Modes (Continued)
key.
or
STOP BOLUS
key.
soft key.
-digit code.
soft key.
soft key as
Alaris
Patient Controlled Analgesia Module Section
A
Summary
Morphine
1
Loading
mg
Dose:
[Conc]: 1 mg/mL
>Press START to Close
Summary
STOP
START
LOAD
A
Guardrails Drug Setup
IMS Pre-fill 30mL
2
PCA
mg
DOSE
LOCKOUT
20
INTERVAL
minutes
CONT
2
mg/h
DOSE
MAX
25
LIMIT
mg/4 h
[Conc]: 1 mg/mL
>Press CONFIRM
CHANGE
CONFIRM
MODE
System Directions for Use
®