Primary Infusion
IV Fluid Infusion (Continued)
Pausing, Restarting, Restoring Infusion
Reference "Continuous Infusion" procedure.
Changing Rate or VTBI During Infusion
Reference "Continuous Infusion" procedure.
NOTES:
Syringe Module: When
option is not available.
The infusion may be paused by pressing the
Reference "Continuous Infusion", "Pausing and Restarting
Infusion" procedure.
To view infusion Time Left during a volume/duration infusion,
press
CHANNEL SELECT
press
soft key.
START
2-38
Programming
- With Guardrails
Suite MX Protection (Continued)
®
is disabled, the
ALL MODE
key. To return to previous screen,
VTBI ALL
soft key.
PAUSE
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Fluid Setup
A
TPN
4.17
mL/h
RATE
ALL
VTBI
00 45
Time Left:
h
min
>Press START
VOLUME
PAUSE
START
DURATION
Alaris
System Directions for Use
®
Pump and Syringe Modules Section