Trumpet and Start-Up Curves (Continued)
Introduction (Continued)
Pump Module
Effects of Pressure Variations
Under conditions of
typically exhibits a long-term accuracy offset of approximately
from mean values.
-0.7%
Under conditions of
typically exhibits a long-term accuracy offset of approximately
from mean values.
-4.2%
Under conditions of
typically exhibits a long-term accuracy offset of approximately
from mean values.
+4.4%
Resulting trumpet observation points typically track those of
accuracy; therefore, no significant change in short-term
variations result under these pressure conditions.
Effects of Negative Solution Container Heights
With a negative head height of
typically exhibits a long-term accuracy offset of approximately
from mean values.
-3.1%
Resulting trumpet observation points typically track those of
accuracy; therefore, no significant change in short-term
variations result under negative head height conditions.
Syringe Module
Trumpet and start-up curves have been provided for
and
mL/h. Measurements for trumpet curve rates below
5.0
mL/h are not provided because of the difficulty in
1.0
measuring extremely small volumes over a large duration of
time. In this case, the linear relationship of the plunger
position and velocity to syringe volume and rate is verified,
and is a function of the accuracy of the design.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
mmHg pressure, the Pump Module
+100
mmHg pressure, the Pump Module
+300
mmHg pressure, the Pump Module
-100
meters, the Pump Module
-0.5
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
mL/h
1.0
2-103
General Information