Administration Sets

WARNINGS
• When priming:
Ensure patient is not connected.
Ensure air is expelled from line prior to beginning
infusion (unexpelled air in line could have serious
consequences).
Failure to prime correctly can delay infusion delivery and
cause the total volume to be infused to read higher than
the actual total delivered to the patient.
• Discard if packaging is not intact or protector caps are
unattached.
• Use only Alaris
Pump Module/Gemini Infusion
®
System administration sets with the Alaris
Module. The use of any other set may cause improper
instrument operation, resulting in an inaccurate fluid
delivery or other potential hazard. For a list of compatible
sets, reference the Set Compatibility Card (provided
separately).
• Use only standard single-use disposable syringes (with
luer-lock connectors) and administration sets, designed
for use on syringe pumps, with the Alaris
Module. The use of any other syringe or
administration set may cause improper instrument
operation, resulting in an inaccurate fluid delivery or
pressure sensing, or other potential hazards. For a list of
compatible syringes, reference "Compatible Syringes". For
a list of compatible sets, reference the Set Compatibility
Card (provided separately).
• Before loading or unloading the syringe, always turn
off fluid flow to the patient, using the tubing clamp or
stopcock. Uncontrolled fluid flow can occur when the
administration set is not clamped or turned off, and
may cause serious injury or death.
2-74
General Information
Warnings and Cautions
(Continued)
FU
Pump
®
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Syringe
FU
®
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Alaris
System Directions for Use
®
Pump and Syringe Modules Section

Advertisement

Table of Contents
loading

Table of Contents