Alarms - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Alert
Clinical Advisory
Audio Characteristics
Type
Switchover
Pump and Syringe Modules
Alarm
Channel Disconnected
2-110
Troubleshooting and Maintenance
Alarms, Errors, Messages (Continued)
Definitions
A visual message to help reduce programming errors by indicating a Limit
("Soft" or "Hard") has been exceeded. A response is required before
programming can continue.
A visual message when a designated drug is selected, to remind clinician of
specific hospital/facility standards of practice when programming an
medication. A specific Clinical Advisory and/or message can be associated
with a selected drug within any of the patient care profiles. Clinical Advisories
will not be displayed in Anesthesia mode.
Six short beeps: secondary
switching to primary. Two short
beeps: bolus switching to
continuous.

Alarms

Module(s) disconnected while in
operation or have a communication
problem.
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Sound
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Meaning
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Notes
Variable volume; can be silenced
and disabled in System
Configuration.
Response
To silence alarm and clear
message from screen, press
soft key. Reattach
CONFIRM
module if desired, ensuring it is
securely "clicked" into place at
Module Release Latch. If alarm is
still present, replace module with
an operational instrument.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
IV