Syringe Module (Continued)
Critical Volume:
Dimensions:
Environmental Conditions:
Temperature Range:
Relative Humidity:
(Avoid prolonged exposure
to relative humidity >85%)
Atmospheric Pressure:
Equipment Orientation:
Flow Rate Programming:
Rate Restriction by Syringe Size:
Fluid Ingress Protection:
2-98
General Information
Specifications (Continued)
Maximum over-infusion which can occur in the event of a single-fault
condition will not exceed
and
of maximum syringe travel after syringe loading.
1%
x
x
4.5"W
15.0"H
Operating
-
41
104°F
(
-
)
5
40°C
-
20
90%
Noncondensing
-
mmHg
525
4560
(
-
hPa)
700
6080
To ensure proper operation, PC Unit must remain in an upright position.
Flow rate range is from
Flow Rates (mL)
-
0.01
9.99
-
10
99.9
-
100
999
Syringe Size (mL)
/
50
60
30
20
10
5
3
1
, Drip Proof
IPX1
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
of nominal syringe fill volume during loading
2%
7.5"D
Storage/Transport
-
-
4
140°F
(-
-
20
60°C
-
5
85%
Noncondensing
-
mmHg
375
760
(
-
500
1013
to
mL/h and can be selected as follows:
0.01
999
Selectable Increments (mL/h)
0.01
0.1
Flow Rate Range (mL/h)
-
0.1
-
0.1
-
0.1
-
0.1
-
0.1
0.01
0.01
)
hPa)
1
999
650
500
250
150
-
100
-
30
Alaris
System Directions for Use
®
Pump and Syringe Modules Section