Programming An Infusion With Pca Pause Protocol Enabled - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Programming an Infusion with PCA Pause Protocol Enabled

1.
Perform steps
1
an Infusion".
2.
Confirm drug and concentration selections and press Yes
soft key.
3.
If facility has chosen to enable optional PCA Pause
Protocol, review Clinical Advisory. To continue, press
soft key.
CONFIRM
To activate PCA Pause Protocol, attach and start an
Alaris
t
Module and/or Alaris
®
E
CO
2
facility protocol.
• Verify appropriate monitoring modules are attached to
PC Unit and press
4.
Verify parameters are correct and press
confirm.
• Clinical Advisory appears.
5.
Press
CONFIRM
6.
Start applicable infusion, as described in following
procedures:
Setting Up PCA Dose Only
Setting Up Continuous Infusion Only
Setting Up PCA Dose + Continuous Infusion
Setting Up Loading Dose Only
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
-
in "Preparing Infusion", "Programming
8
®
soft key.
CONFIRM
soft key.
p
Module per
S
O
2
soft key to
NEXT
A
Clinical Advisory:
Attach an SpO2 or EtCO2
module now.
The drug selected will
activate the PCA Pause
Protocol once a monitoring
device is operating.
>Press CONFIRM
EXIT
CONFIRM
A
PCA Pause Limits
should be reviewed.
Press PAUSE LIMITS to
review the current settings.
Press CONFIRM to continue
programming the PCA
infusion.
>Press CONFIRM
PAUSE
CONFIRM
LIMITS
Programming
3-33

Advertisement

Table of Contents
loading

Table of Contents