Syringe Module - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in todayā€™s growing professional healthcare environment
Table of Contents

Advertisement

Pump Module (Continued)
Time to Alarm, Maximum:
Volume to be Infused
Programming Increments:
Weight:
Bolus Volume following
Occlusion (at intermediate rate):
Without Pressure Sensing Disc:
With Pressure Sensing Disc:
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Specifications (Continued)
Rate (mL/h)
1
25
Range
(mL)
0.1 - 9.99
10 - 999.9
1000 - 9999
2.5 lbs

Syringe Module

Pressure Setting
Low
Medium
High
mmHg
300
mmHg
500
mmHg
1000
Alaris
System has a back-off safety feature which, when enabled and a
®
pressure sensing disc is in use, is designed to reduce bolus volume on
occlusion release.
WARNING
Installing a pressure sensing disc after an infusion has started can
result in a bolus to the patient.
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pressure Limit (mmHg)
50
≤5 minutes
≤15 seconds
Increments
(mL)
0.01
0.1
1
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Back Off Disabled
0.277
0.416
0.764
525
≤45 minutes
≤2 minutes
Bolus (g)
0.329
0.523
0.736
Back Off Enabled
0.098
0.136
0.137
General Information
2-97

Advertisement

Table of Contents
loading

Table of Contents