Messages - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Message
Battery Run Time = X.X hours
Low Battery
Panel Locked
Panel Unlocked
Powering Down
Replace Battery
Plug the PC Unit into an
fully charged battery when needed.
be on whenever the PC Unit is plugged in.
Alaris
System Directions for Use
®
PC Unit Section
Alarms, Errors, Messages (Continued)

Messages

power cord is disconnected
AC
from power source. Approximate
remaining battery run time under
current power consumption rate is
displayed.
Low battery threshold sensed;
remaining battery run time is
limited.
Tamper Resist feature is active and
key was pressed.
Tamper Resist feature deactivated.
Last module powering off. System
shuts off in indicated number of
seconds.
Occurs at System On. Battery has
less than
Storage
outlet during storage, to ensure a
AC
indicator light (
AC
Meaning
of original capacity.
50%
) will
Response
None. Connect
AC
power source as soon as possible.
Connect
power cord to power
AC
source; alarm will be silenced.
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
Press any key, except
key, to cancel power down
sequence.
Press either
SYSTEM OFF
soft key to continue
CONFIRM
normal operation with reduced
battery capacity. Service by
qualified personnel is required.
Troubleshooting and Maintenance
power cord to
SYSTEM ON
or
1-39