Accumulated Air Window:
Bolus Volume following
Occlusion, Maximum:
Critical Volume:
Dimensions:
Environmental Conditions:
Atmospheric Pressure:
Relative Humidity:
(Avoid prolonged exposure
to relative humidity >85%)
Temperature Range:
Equipment Orientation:
Electrical Classification:
Flow Rate Programming
Increments:
Fluid Ingress Protection:
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Specifications
Pump Module
Single Bolus
Setting
50
75
250
*500
* In Anesthesia Mode only.
Rate (mL/h)
25
The maximum over-infusion which can occur in the event of a single fault
condition is
mL.
0.6
x
x
3.3"W
8.9"H
5.5"D
Operating
-
mmHg
525
4560
(
-
700
6080
-
20
90%
Noncondensing
-
41
104°F
(
-
)
5
40°C
To ensure proper operation, the system must remain in an upright position.
Class
, Type
1
CF
Rate Range
(mL/h)
0.1 - 9.99
10 - 99.9
100 - 999
, Drip Proof
IPX1
CHANNEL
Volume Window
(mL)
2.8
8.0
8.0
12.0
Pressure Limit (mmHg)
50
≤0.3 mL
Storage/Transport
375
hPa)
(
500
5
Noncondensing
-
4
(-
20
Defibrillator Proof
Increments (mL/h)
User Input Rates
Device Calculated Rates
0.1
1
FU
CHANNEL
SELECT
PAUSE
OFF
RESTART
% Air that Causes
Alarm
10%
20%
30%
30%
525
≤0.6 mL
-
mmHg
760
-
hPa)
1013
-
85%
-
140°F
-
)
60°C
0.01
0.1
1
General Information
2-95