Cardinal Health Alaris 8000 Series Directions For Use Manual page 156

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Syringe Module (Continued)
Time to Alarm, Maximum:
Volume to be Infused
Programming Increments:
Weight:
NOTES:
On a high setting, the actual occlusion pressure will vary based on the syringe size and manufacturer.
Flow rates as low as
limit adjusts to
0.1
The Maximum Time to Alarm specifications are based on Cardinal Health's standard operating conditions:
Atmospheric Pressure:
Back Pressure:
Humidity:
Temperature:
Syringe Type:
2-100
General Information
Specifications (Continued)
Rate (mL/h)
Range (mL)
-
0.1
9.99
-
10
60
4.5 lbs
mL/h are available only with
0.01
mL/h.
-
mmHg
645
795
mmHg before producing occlusion
0
-
20
90%
68 ±4°F
/
cc
BD 50
60
High Setting
1
120 minutes
5
30 minutes
Increments (mL)
0.01
0.1
cc and
1
3
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pressure Limit
No Disc
With Disc Highest
(1000 mmHg) Setting
cc syringes. For larger syringes, lower
Alaris
System Directions for Use
®