Cardinal Health Alaris 8000 Series Directions For Use Manual page 80

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Primary Infusion
Continuous Infusion (Continued)
• If programmed continuous dose infusion is outside
Guardrails
Soft Limit ("Soft Limit") for that care area, a
®
prompt appears before programming can continue. If
Yes soft key is pressed, programming continues; if No
soft key is pressed, infusion needs to be
reprogrammed.
If programmed continuous dose infusion is outside
Guardrails
Hard Limit ("Hard Limit") for that care area,
®
a prompt appears before programming can continue.
Infusion needs to be reprogrammed.
If a dose outside of Soft Limits has been entered and
verified as correct, Message Display also shows either
"
" for a low dose or "
LLL
Pausing and Restarting Infusion
1.
Press
key.
PAUSE
scrolls in Message Display.
PAUSE
appears on Main Display.
PAUSED
• Yellow Standby Status Indicator illuminates.
• After
minutes,
2
audio prompts begin, and yellow Standby Status
Indicator flashes.
2-24
Programming
- With Guardrails
" for a high dose.
↑↑↑
PAUSE-RESTART CHANNEL
Suite MX Protection (Continued)
®
visual and
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A
Alteplase
Dose exceeds
Yes
Guardrail limit of
No
.81 mg/kg/h.
Proceed?
>Press Yes or No
Guardrails Drug Setup
A
Alteplase
Dose exceeds
Reprogram
Guardrails hard
limit of .82 mg/kg/h.
>Press REPROGRAM
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Alaris
System Directions for Use
®
Pump and Syringe Modules Section