General Information; Warnings And Cautions - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

WARNINGS
• The nurse call feature should not be used as the primary
source of alarm notification. The audible and visual
alarms of the Alaris
clinical signs and symptoms, are the primary sources
providing notification that an alarm condition exists.
• Before each use, ensure the nurse call cable is fully
inserted into the jack, and test the operation of the nurse
call system by causing an alarm and verifying proper nurse
call operation.
CAUTIONS
• To assure compliance with Electromagnetic
Compatibility Standards (
option must only be used with a properly configured
hospital interface cable assembly. The approved hospital
nurse call interface cable consists of an
cable with an attached Fair-Rite Suppressor Core
24–660
Assembly,
P/N 0443167251
with the Nurse Call Accessory. Refer to the cable
assembly instructions, included with the Nurse Call
Accessory, for additional information.
• When moving the PC Unit, ensure the nurse call cable is
disconnected from the nurse call jack on the PC Unit and
the fluid seal covers the jack opening.
• Carefully route cabling to reduce the possibility of
entanglement that could result in a tripping hazard.
Alaris
System Directions for Use
®
Nurse Call Accessory Section

Warnings and Cautions

System, used in conjunction with
®
), the nurse call
IEC 60601-1-2
MCM
. The Fair-Rite part is included

General Information

Electronics
7-3
General Information