Primary Infusion
Continuous Infusion (Continued)
• If Yes was selected and facility has defined a
Guardrails
Clinical Advisory ("Clinical Advisory") for
®
that drug, a message appears. To indicate information
has been noted and continue programming, press
soft key.
CONFIRM
• If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
• If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
• If selected drug is not weight-based, Not Used
displays in
PATIENT WEIGHT
• If hospital/facility practice guidelines identify selected
drug as weight-based, prompt for a patient weight in
kilograms appears (as in illustrated example, which
reflects use of Alteplase).
4.
Verify parameters are correct and press
confirm.
• Syringe Module: If
displays.
2-22
Programming
- With Guardrails
Suite MX Protection (Continued)
®
field.
NEXT
Mode is enabled,
ALL
soft key to
VTBI ALL
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A
Alteplase
Clinical Advisory:
This dosing is for Acute
Ischemic STROKE
>Press CONFIRM
CONFIRM
A
Guardrails Drug Setup
Alteplase
100
DRUG
mg
AMOUNT
100
DILUENT
mL
VOLUME
PATIENT
Used
WEIGHT
TIME
hour
UNITS
mg/kg/h
DOSING
UNITS
[Conc]: 1 mg/mL
>Press NEXT to Confirm
DRUG
NEXT
LIBRARY
A
Guardrails Drug Setup
Alteplase
100
DRUG
mg
AMOUNT
100
DILUENT
mL
VOLUME
_ _ _ _
PATIENT
kg
WEIGHT
TIME
hour
UNITS
mg/kg/h
DOSING
UNITS
[Conc]: 1 mg/mL
>Enter Patient Weight
DRUG
LIBRARY
Alaris
System Directions for Use
®