Secondary Infusion
Infusion (Continued)
7.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
To enter duration, press
•
numeric data entry keys (rate value is calculated and
displayed).
• To enter rate, press
numeric data entry keys.
8.
Verify parameters are correct and press
• If programmed duration is outside Soft Limit for that
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
If programmed duration is outside Hard Limit for that
•
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
Pausing, Restarting, Restoring Infusion
Reference "Primary Infusion - With Guardrails
Protection", "Continuous Infusion" procedure.
Changing Rate or VTBI During Infusion
Reference "Primary Infusion - With Guardrails
Protection", "Continuous Infusion" procedure.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
- With Guardrails
(
) (Continued)
Pump Module
soft key and use
DURATION
soft key and use
RATE VOLUME
START
Suite MX
®
Suite MX
®
Suite MX Protection
®
soft key.
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Guardrails Drug Setup
A
Methotrexate
SECONDARY
4
mL/h
RATE =
ALL
VTBI
6:00
DURATION
hh:mm
1200
DOSE =
mg/m
2
[Conc]: 342 mg/mL
>Verify Secondary Clamp
Open, Then Press START
RATE
PAUSE
SETUP
START
VOLUME
Programming
2-43