Press and hold
CHANNEL OFF
approximately
seconds.
1.5
NOTE:
If no other channel is active, the system powers down when
the
CHANNEL OFF
Selecting Pressure Limit
1.
Press
CHANNEL SELECT
2.
Press
OPTIONS
3. Press Pressure Limit soft key.
4.
To select a pressure limit, press appropriate soft key.
5.
Press
CONFIRM
NOTE:
Option to change pressure limit can be selected:
•
after drug is selected, and before infusion mode is
selected and infusion starts, or
•
after infusion starts.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Infusion Modes (Continued)
Stopping Infusion
key until a beep is heard,
key is released.
key.
key.
soft key.
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Channel Options
2 of 2
Pressure Limit - High
Prime Set with Syringe
>Select an Option or
EXIT
PAGE
EXIT
UP
A
Pressure Limit Selection
Low
Med
High
CONFIRM
Programming
3-31