Cardinal Health Alaris 8000 Series Directions For Use Manual page 70

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Preparing Syringe and Administration Set (
Priming (continued)
Pressure Sensing Disc Installed (Continued)
6.
Reinstall pressure sensing disc, as follows:
a. Orient pressure sensing disc, as follows:
• fluid side up (patient side down)
• cavity forward (membrane toward instrument)
b. Gently slide pressure sensing disc up into slot in
pressure sensing disc housing.
c. Apply firm upward pressure on pressure sensing disc
(not tubing) until disc snaps into place.
7.
To return to main screen, press
• If
soft key is pressed before pressure sensing
EXIT
disc is reinstalled, a prompt to reinstall pressure
sensing disc displays.
2-14
Getting Started
Syringe Module
soft key.
EXIT
FU
RATE (mL/h)
)
(Continued)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Syringe Module
Slot in Pressure
Fluid Side
Sensing Disc
Housing
Pressure
Sensing Disc
Patient Side
Pressure Sensing Disc
A A A
Re-install Pressure Sensing
Disc.
>Press CONFIRM
CONFIRM
Alaris
System Directions for Use
®
Pump and Syringe Modules Section