Pump Module (Continued)
Occlusion Pressure
Secondary Infusions
Syringe Module
All Mode
Auto Pressure
Auto Pressure Limit Adjustment
2-84
General Information
Features and Displays (Continued)
A complete range of downstream occlusion detection options is
provided.
• Pump mode: Downstream occlusion alarm threshold is
mmHg at flow rates of
525
mL/h, occlusion pressure is rate-dependent, to ensure rapid
<30
response to occlusions.
• Selectable pressure mode: Downstream occlusion alarm
threshold can be adjusted in
maximum occlusion pressure of
• Auto-Restart: (See "Auto-Restart" definition.)
In addition, Alaris
Dual rate sequential piggyback (secondary) infusions may be
infused, with limits, at delivery rates and volumes independent of
primary infusion parameters. Automatic changeover occurs to
primary infusion parameters when secondary infusion is complete if
an Alaris
administration set is used.
When
contents of the syringe will be delivered.
When enabled and a pressure sensing disc is in use, Auto
Pressure option is displayed in Pressure Limit screen. Auto
Pressure automatically sets alarm limit for a shorter time to alarm,
as follows:
• If current pressure is
30
• If current pressure is greater than
30%
When a bolus is delivered, pressure alarm limits are temporarily
raised to maximum limit.
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
System provides fluid-side occlusion detection.
®
Pump Module/Gemini Infusion System check valve
®
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
is selected as the volume to be infused (
ALL
100
mmHg to current pressure, to create a new alarm limit.
to current pressure, to create a new alarm limit.
mL/h or greater. For rates
30
mmHg increments, up to
25
mmHg.
525
mmHg or less, system adds
mmHg, system adds
100
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
), entire
VTBI