Do you have a question about the Kangaroo Omni 385400 and is the answer not in the manual?
Questions and answers
Jacqueline Nunnally
January 20, 2025
Why does the alarm for blockage on patient side keep going off even though there is no blockage. I have changed the set,I have cleaned the machine and every 5 minutes it is going off again
2 comments:
Mr. Anderson
February 4, 2025
The blockage alarm on the Cardinal Health Kangaroo Omni 385400 could be caused by an issue with the occlusion sensor region being dirty or malfunctioning. To address this, remove the cassette, clean and dry the occlusion sensor region, and reload the cassette. If the alarm continues, the sensor may still be detecting an issue despite no actual blockage.
This answer is automatically generated
Yazz
March 25, 2025
Mine does the same thing, the machine isn't even 2 weeks old. You can leave it running with the end unattached and see it's running fine and sometimes it just keeps sounding the alarm every couple of minutes. You can even **** on it like a straw so the sensor should be reading it put out faster than intended, still says it (lol). I increased the flush frquency to every hour and reduced the amount proportionally and it happens much less often now. Just for reference mine is being loaded with Kate Farms 1.5 peptide.
Zoey
April 23, 2025
How to change feed time on kangaroo omni
1 comments:
Mr. Anderson
May 12, 2025
To change the feed time on the Cardinal Health Kangaroo Omni 385400, follow these steps:
1. Go to the appropriate feed setup screen. 2. Use the controls to adjust the time interval for feeding. 3. Press the OK button to confirm the time interval and return to the Adjust Flush screen. 4. Press the Done button to confirm and return to the Ready to Feed screen.
Adjustments can only be made before starting the feed. If feeding has already started, press the Stop button to make changes.
This answer is automatically generated
Willie Haygood
March 5, 2025
How to clear volume on kangaroo Omni pump
1 comments:
Mr. Anderson
May 14, 2025
To clear volume on the Cardinal Health Kangaroo Omni 385400 pump:
1. Go to the More Options screen. 2. Use the arrow keys to select "Clear Total Volumes" and press the Select button. 3. Press the Clear button. 4. Press the Back button to return to More Options.
This answer is automatically generated
Patti Bishop
February 18, 2025
When using Auto Prime the formula exits the cassette filled with air bubbles. Removed cassette and cleaned area. No resolution of problems. No alarms
1 comments:
Mr. Anderson
May 14, 2025
Air bubbles in the cassette when using Auto Prime with the Cardinal Health Omni 385400 could be caused by air trapped in the bag or tubing before priming. This may happen if the feeding bag is not properly filled or if the tubing is not fully primed, allowing air to enter the system.
This answer is automatically generated
Virginia and James Goodwin
February 4, 2025
After priming the screen with start is not appearing
1 comments:
Mr. Anderson
February 10, 2025
The start screen may not be appearing after priming the Cardinal Health Kangaroo Omni 385400 due to an incomplete priming process or an issue with the pump setup. Ensure that:
1. The feeding solution has exited the distal connector before pressing "Done." 2. The patient is disconnected during priming. 3. The cassette is properly loaded in the pump. 4. There are no alarms, such as "Feed Bag Empty" or "Patient Tube Blocked."
If the issue persists, try reloading the cassette and ensuring all steps are completed correctly.
This answer is automatically generated
Harold Lodevil
June 14, 2025
How do I set language back to english
Teresa Brennan
May 23, 2025
I am wondering more about the wifi connection possibilities
Mona Buffum
April 30, 2025
can the pump be set to flush 35 ml every hour
Related Manuals for Cardinal Health Kangaroo Omni 385400
Glossary Definitions Accuracy: The difference between the prescribed hourly feed rate and the amount of enteral fluids actually delivered. Continuous Feeding: Feeding mode where a patient is fed continuously at a prescribed flow rate. Enteral Fluids: Fluids like water and available formulas that are used for enteral nutrition administration via the human gastrointestinal tract.
Page 5
Intermittent Feeding: Feeding mode where a patient is fed a predetermined amount of enteral fluids at a prescribed flow rate over specified intervals throughout the day. Rotor: The black circular wheel on Kangaroo OMNI™ Enteral Feeding Pump that rotates to push fluid through the Kangaroo OMNI™...
Glossary Symbols SYMBOL DEFINITION SYMBOL DEFINITION SYMBOL DEFINITION Authorized representative in the Non-Sterile Keep dry European Community Not made with natural China RoHS (reduction of Catalog number rubber latex hazardous substances) Federal (USA) law restricts Kangaroo OMNI™ Enteral Feeding Dispose of as Electrical and Manufacturer Pump to sale by or on the order of Electronic Waste...
Introduction Intended Purpose (Intended Use) The Kangaroo OMNI™ Enteral Feeding Pump with Kangaroo OMNI™ Feeding Sets are intended for delivery of enteral fluids, including nutritional fluids and/or water to the gastrointestinal system via nasogastric, orogastric, nasojejunal, gastrostomy, and jejunostomy tubes. Not for use with neonates. Indications for Use The Kangaroo OMNI™...
Introduction This enteral feeding pump contains an electronic circuit board. When discarding the pump at the end of its useful life, it is recommended to remove the back label, unscrew the back housing from the pump, and remove the circuit board from the pump.
14. Do Not use Kangaroo OMNI™ Enteral Feeding Pump adjacent to or stacked with other equipment because it could result in improper operation. If such use is necessary, Kangaroo OMNI™ Enteral Feeding Pump and the other equipment should be observed to verify that they are operating normally. 15.
Page 10
17. The power adapter cord, Kangaroo OMNI™ Feeding Set tubing, and Kangaroo OMNI™ Enteral Feeding Pump accessories may cause a tripping hazard. Avoid leaving wires, cords, or tubing in a pathway where a person could trip on them and sustain injury. 18.
Considerations for Infants and Children 1. Thick formula is not recommended for use in infants <6 months of age. 2. Use caution when administering thick formula in patients who cannot tolerate delivery of the prescribed volume within 10% accuracy. This includes, but is not limited to infants, critically ill patients or those who may experience a clinical impact due to underfeeding or overfeeding.
Battery Pack Caution: The battery cells used in Kangaroo OMNI™ Enteral Feeding Pump may present a fire or chemical hazard if mistreated. Do not disassemble, heat above 60ºC (140ºF), or incinerate. Avoid exposing the battery pack to heat or fire since a fire or explosion hazard could result.
Kangaroo OMNI™ Enteral Feeding Pump Overview Kangaroo OMNI™ Enteral Feeding Pump Overview: Front Selection Buttons: Press the selection buttons next to the desired option on the screen. Night Mode Button: Press once to switch screen to night mode. Press again to return to normal mode. Power Indicator Light: Green light confirms Kangaroo OMNI™...
Kangaroo OMNI™ Enteral Feeding Pump Overview Kangaroo OMNI™ Enteral Feeding Pump Overview: User Interface Title Bar: Communicates current running status of Kangaroo OMNI™ Enteral Feeding Pump. Flow Indicator: Moving droplet indicates that Kangaroo OMNI™ Enteral Feeding Pump is delivering enteral fluid and is working normally.
Kangaroo OMNI™ Enteral Feeding Pump Overview Kangaroo OMNI™ Feeding Set Overview The below is an example of a Kangaroo OMNI™ ENtelliSet™ Feeding Set with Flush Bag for illustration purposes only. Note: product availability and type vary by region. Flush Bag: (when applicable) intended for use in the administration of water to help manage patient hydration requirements and help keep the feeding tube patent to avoid tube clogs.
Page 16
Kangaroo OMNI™ Enteral Feeding Pump Overview Kangaroo OMNI™ Feeding Set Overview CASSETTE COLOR PRODUCT NAME DESCRIPTION Feeding set designed to deliver available formula (categorized as Clear Kangaroo OMNI™ Feeding Set level 0-1 within the IDDSI framework) Feeding set designed to deliver available formula Kangaroo OMNI™...
Kangaroo OMNI™ Enteral Feeding Pump Overview Screen Color Meaning Blue Background: Green Background: Brown Background: Configuring pump for use Pump running Pumping paused Feeding set valve closed Droplet will be moving Droplet will be paused Feeding set valve open Feeding set valve open Orange Background: Light blue Background: Yellow Background:...
System Setup Important Note: A newly purchased Kangaroo OMNI™ Enteral Feeding Pump comes with Kangaroo OMNI™ Enteral Feeding Pump, AC power adapter, Kangaroo OMNI™ Feeding Set, and pole clamp. Kangaroo OMNI™ Enteral Feeding Pump and along with Kangaroo OMNI™ Feeding Sets and accessories, need to be installed and put into service in accordance with the information provided by this manual and accompanying documents.
System Setup 2. Attach the Pole Clamp to Kangaroo OMNI™ Enteral Feeding Pump A. Use the screw provided to attach the pole clamp to the back of Kangaroo OMNI™ Enteral Feeding Pump. The pole clamp can be attached in a horizontal or vertical orientation. Optional: Place the power cable into the channel on the back of Kangaroo OMNI™...
System Setup 2b. Clear previous rate/feed settings: A. Press the Clear Settings button. B. Press the Adjust Rate button or the Adjust Flush button and make any desired adjustments. (See Set Feed/Rate section below for additional information). Hang Feeding Bag/Feeding Container on IV Pole CAUTION: The top of the starting volume of enteral fluids should be 10 inches above the top of Kangaroo OMNI™...
System Setup Prime Kangaroo OMNI™ Enteral Feeding Pump Note: If the Kangaroo OMNI™ Feeding Set is already primed, press the Skip Prime button. 5a. Auto Prime Kangaroo OMNI™ Enteral Feeding Pump Feed Only or Feed with Flush Feeding Set To prime the Kangaroo OMNI™ Feeding Set, first fill the bag(s) with feeding solution or connect a pre-filled container.
Page 23
System Setup 5b. Manually Prime Kangaroo OMNI™ Feed Only Feed Set To prime the Kangaroo OMNI™ Feeding Set, first fill the bag(s) with feeding solution or connect a pre-filled container. A. Press Prime. WARNING: Patient must not be connected to Kangaroo OMNI™ Enteral Feeding Pump when priming is performed.
System Setup B. Press Manual Prime. C. Press and hold the Prime Feed or Prime Flush button until the solution just begins to exit the distal connector. Note: When using Manual Prime feature; Feed Bag Empty, and Patient Tube Blocked alarms are disabled. Flush options will only be present when a feeding set with flush bag is used.
Page 25
System Setup 6a. Adjusting Continuous Feeds Continuous Feeds is a feeding mode where a patient is fed continuously at a prescribed flow rate. The Rate feature on Kangaroo OMNI™ Enteral Feeding Pump is used to establish a fixed volume per hour (mL/hr) rate that will continue until the feed runs out or Kangaroo OMNI™...
System Setup To Enable Intermittent Mode: A. Press the More Options button on the Ready to Feed screen. B. Press the down arrow button until Intermittent Mode is highlighted. C. Press the Select button. "On" or "Off" will be displayed to indicate the present Intermittent Mode state.
Page 27
System Setup C. Press the + and – buttons to set the number of feeds to be delivered. D. Press OK to save the settings and return to the Adjust Feed screen. E. Press the Adjust Volume button to set the Feed volume. F.
Page 28
System Setup J. Press OK to save the settings and return to the Adjust Feed screen. K. Press Done to return to the Ready to Feed screen. 6d. Adjusting Flush (Feed with Flush Kangaroo OMNI™ Sets) A. Press the Adjust Flush button from Ready to Feed screen. B.
System Setup D. Press the OK button to confirm the flush volume and return to the Adjust Flush screen. E. Press the Adjust Interval button. F. Press the + and – buttons to input the desired time interval. G. Press the OK button to confirm the time interval and return to the Adjust Flush screen.
System Setup Intermittent Mode Continuous Mode B. To change the rate or feed setting once feeding has started, press the Stop button. Note: A green screen with a moving droplet means the feeding is in progress. Intermittent Mode Turn Off Feed Complete Audio (Optional) A feeding regime will end with either a "Feed Error"...
System Setup C. Press the Select button to toggle the audio between "On" and "Off". D. Press the Done button to return to the Ready to Feed screen. Feed Complete Information Signal Pump Status Indicator: blinking green The FEED COMPLETE screen appears at the end of a single or multiple feed. A.
System Setup B. Press Restart Feeds if you are beginning a new, full feeding regimen. Turn Off Kangaroo OMNI™ Enteral Feeding Pump A. Press and hold Power Button for 3 seconds. Note: Turning off Kangaroo OMNI™ Enteral Feeding Pump will save settings.
Additional Options ADDITIONAL OPTIONS WHILE FEEDING IS IN PROGRESS Stop Feeding Press the Stop button to end the feeding at any time. Flush Now While Kangaroo OMNI™ Enteral Feeding Pump is running with a feed and flush set installed, Kangaroo OMNI™ Enteral A.
Additional Options Pause Feeding The Timed Pause feature allows Kangaroo OMNI™ Enteral Feeding Pump to be paused temporarily for a period of 5 to 240 minutes. While paused, Kangaroo OMNI™ Enteral Feeding Pump will move a very small amount of feeding solution through the Kangaroo OMNI™...
Additional Options Lock / Unlock Input Screen Kangaroo OMNI™ Enteral Feeding Pump allows the user to lock the input screen to prevent accidental button presses during portable use. This is especially helpful when transporting Kangaroo OMNI™ Enteral Feeding Pump in a backpack. Lock Input Screen A.
Additional Options D. Press the Done button to return to Ready to Feed. Clear Total Volume: This feature resets the feeding and flushing counters to zero. Use for starting a new patient or a new feeding. A. In the More Options screen, use the arrow keys to select Clear Total Volumes, and press the Select button.
Additional Options View History: This feature retrieves Kangaroo OMNI™ Enteral Feeding Pump history for the last 30 days. A. In the More Options screen, use the arrow keys to select View History, and press the Select button. B. Press the View Hours button or the View Days button. C.
Additional Options C. Press the OK button to return to More Options. Adjust Brightness: This feature adjusts the brightness of the display backlight. A. In the More Options screen, use the arrow keys to select Adjust Brightness, and press the Select button. B.
Cap Storage and Pole Clamp Setup Using the ENFit™ Cap Storage Feature A. Remove cap from tube connector. B. Secure cap while not in use. C. Replace cap on tube connector after disconnecting from patient. Attach Adjustable Pole Clamp to IV Pole CAUTION: When attaching Kangaroo OMNI™...
Biotech Mode Biotech Mode Settings Biotech Mode is a menu location for certain Kangaroo OMNI™ Enteral Feeding Pump settings that are generally only changed by a biomedical technician. Settings in Biotech Mode include Lock Settings, Interruption Monitor, Timed Pause, and Language. Access Biotech Mode A.
Biotech Mode C. Press the Select button to activate the highlighted feature, and then press Done. D. Press the Restart Pump button to save the settings and exit Biotech mode. Lock Settings The Lock Settings feature disables the user input of Kangaroo OMNI™ Enteral Feeding Pump settings, allowing clinicians to ensure that settings do not change.
Biotech Mode C. Press the Select button. D. Press the Done button. E. Press Restart Pump button to save the settings and exit Biotech mode. If a user attempts to modify any settings, a Settings Locked reminder screen will be displayed. The Lock Settings feature can only be disabled by entering Biotech Mode and switching Lock Settings to Off.
Page 43
Biotech Mode C. Press the Select button. D. Use the arrow keys to select the desired language. E. Press the Select button. A check mark should appear in the check box next to the desired language. F. Press the Back button. G.
Biotech Mode Interruption Monitor Settings A. Press View Options button. B. Use the arrow buttons to highlight Interruption Monitor. C. Press the Select button to toggle between On or Off. D. Press the Done button. E. Press the Restart Pump button to save the settings and exit Biotech mode.
Biotech Mode B. Use the arrow buttons to highlight Timed Pause. C. Press the Select button to toggle between On or Off. D. Press the Done button. E. Press Restart Pump button to exit Biotech mode and return to normal operating mode.
Page 46
Biotech Mode B. Press the Start button. C. Follow the screen prompts as Kangaroo OMNI™ Enteral Feeding Pump tests its operating functions. Wait for the testing to complete. D. "Test Passed" screen will appear when the performance test is complete. Note: Contact Customer Service if you do not obtain a Test Passed result.
Cleaning Caution: Kangaroo OMNI™ Enteral Feeding Pump, and power adapters are not designed to be immersed underwater. Do not immerse Kangaroo OMNI™ Enteral Feeding Pump, or power adapters in water or other cleaning solutions. Failure to follow the cleaning procedures described herein could result in hazards to users, patients, and clinicians. As with any A/C powered electrical device, care must be taken to prevent liquid from entering Kangaroo OMNI™...
Cleaning AC Adapter Cleaning • Refer to General Cleaning Directions before starting. • The AC adapter normally does not require cleaning. If soiling of the AC adapter is observed, unplug from outlet and wipe the exterior surfaces of the AC adapter with a non-abrasive dry or slightly damp cloth. •...
Alarms & Troubleshooting LED Pump Alarm Indicator Lights Kangaroo OMNI™ Enteral Feeding Pump status LED Indicator lights on the upper right of the pump screen provide a quick visual indication of Kangaroo OMNI™ Enteral Feeding Pump status. Solid Green = Pump Running Blinking Green = Ready Solid Yellow = Information Blinking Yellow = Notification...
Alarms & Troubleshooting Verifying Functionality of the Alarm System A. Load a new Kangaroo OMNI™ Feeding Set onto Kangaroo OMNI™ Enteral Feeding Pump. B. Leave the Kangaroo OMNI™ Feeding Set empty. C. Run Auto Prime. D. Upon priming, Kangaroo OMNI™ Enteral Feeding Pump will detect a Feed Error condition and alarm.
Alarms & Troubleshooting Pump Inactive Alarm (Low Priority) LED Pump Alarm Indicator Lights: Solid Yellow PUMP INACTIVE error screen appears if Kangaroo OMNI™ Enteral Feeding Pump has been without input for more than 10 minutes. A. Press Continue to return to the previous screen. Low Battery Alarm Notice (Low Priority) LED Pump Alarm Indicator Lights: Solid Yellow The LOW BATTERY alarm appears when approximately 30 minutes of...
Alarms & Troubleshooting Feed Error Alarm (Medium Priority) LED Pump Alarm Indicator Lights: Blinking Yellow The FEED ERROR screen appears when the feed bag is empty or there is a clog, kink or excessive foam/bubbles in the tubing between the feed bag and the pump.
Page 53
Alarms & Troubleshooting C. Once the air bubbles are out of the occlusion sensor region, restart feeding. D. If alarm continues, turn off pump and Remove the cassette from Kangaroo OMNI™ Enteral Feeding Pump. E. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump occlusion sensor region.
Alarms & Troubleshooting Flush Error Alarm (Medium Priority) LED Pump Alarm Indicator Lights: Blinking Yellow The FLUSH ERROR screen appears when the flush bag is empty or there is a clog or a kink in the tubing between the flush bag and the pump. Detection of this alarm condition may take up to 1 minute to occur.
Alarms & Troubleshooting F. If alarm continues, replace Kangaroo OMNI™ Feeding Set. G. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press Power Off to stop operation and use a different Kangaroo OMNI™ Enteral Feeding Pump. Rotor Stuck Alarm (Medium Priority) LED Pump Alarm Indicator Lights: Blinking Yellow The ROTOR STUCK alarm appears when Kangaroo OMNI™...
Alarms & Troubleshooting Patient Tube Blocked Alarm (Medium Priority) Pump Status Indicator: Blinking Yellow The PATIENT TUBE BLOCKED alarm appears when the feeding solution is no longer being delivered because of a clog between Kangaroo OMNI™ Enteral Feeding Pump and the patient. Detection of this alarm condition may take up to 40 minutes to occur at 1 mL/hr.
Page 57
Alarms & Troubleshooting E. Reload the cassette. F. Press Continue and restart feeding. G. If alarm continues, Replace Kangaroo OMNI™ Feeding Set. H. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press Power Off to stop operation and use a different Kangaroo OMNI™ Enteral Feeding Pump.
Alarms & Troubleshooting Cassette Dislodged Alarm (Medium Priority) Pump Status Indicator: Blinking Yellow The CASSETTE DISLODGED alarm appears when the cassette is not properly loaded in Kangaroo OMNI™ Enteral Feeding Pump. Detection of this alarm condition may take up to 1 second to occur at all flow rates.
Alarms & Troubleshooting B. After correcting irregularities, Reload the cassette. C. If alarm continues, Replace Kangaroo OMNI™ Feeding Set. D. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press Power Off to stop operation and use a different Kangaroo OMNI™ Enteral Feeding Pump.
Alarms & Troubleshooting C. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump valve actuator. D. Reload the cassette. E. If alarm continues, Replace Kangaroo OMNI™ Feeding Set. F. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press Power Off to stop operation and use a different Kangaroo OMNI™ Enteral Feeding Pump.
Page 61
Alarms & Troubleshooting B. Check the tubing inside the cassette to ensure that it is free from lint or other debris. C. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump occlusion sensor region. For excessive cleaning needs, avoid use of sharp objects during the cleaning process.
Alarms & Troubleshooting Sensor Error 2 or 3 Alarm (Medium Priority) Pump Status Indicator: Blinking Yellow The SENSOR ERROR 2 or 3 alarm appears when Kangaroo OMNI™ Enteral Feeding Pump detects an issue with its force sensor. Detection of this alarm condition will occur immediately on power up or cassette removal.
Alarms & Troubleshooting F. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press Power Off to stop operation and use a different Kangaroo OMNI™ Enteral Feeding Pump. Dead Battery Error Alarm (High Priority) Pump Status Indicator: Blinking Red A DEAD BATTERY WARNING will appear when failure of the battery is imminent.
Page 64
Alarms & Troubleshooting B. Press Power OFF. C. Contact Customer Service and provide the three number code displayed on the screen.
Specifications Medical Equipment Kangaroo OMNI™ Enteral Feeding Pump (1) Classified with respect to electrical shock, fire, and mechanical hazards in accordance with ES60601-1 (3rd edition) and UL60601-1 (2nd edition). (2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance with CAN/ CSA C22.2 No.
Page 66
Specifications Note: The amount of water collected in 30 minutes will be half of the value programmed for the hourly flow rate, within ± 10%, or 1.0 mL, whichever is larger. Rate Set Amount Expected in 30 Minutes: 125 mL/hr 56.25 mL – 68.75 mL Flush Accuracy (all Kangaroo OMNI™...
Page 67
Specifications Ambient Air Pressure Transport, storage, and operating atmospheric pressure range from 62 kPA to 106 kPA. Maximum Altitude The maximum altitude for using Kangaroo OMNI™ Enteral Feeding Pump is 4000 m. Be sure to meet ambient air pressure and battery charging limits stated in this manual. The Maximum Infusion pressure is: 25 psi (172 KPa) Type of Protection against Electrical Shock Class II, Internally-powered Equipment...
Customer Service The circuitry of Kangaroo OMNI™ Enteral Feeding Pump should not be repaired or serviced by a customer. Electronic assembly rework by unauthorized technicians can affect accuracy or alarm operation and void the product warranty. All service personnel must be properly trained and qualified with operation of Kangaroo OMNI™ Enteral Feeding Pump.
2 years, or as recommended by facility protocol. This testing may be performed by a qualified Biomedical Engineering Department, an outside service, or by Cardinal Health Factory Service. If the pump malfunctions during use or fails the manual certification test, please contact your Cardinal Health representative or call Customer Service for further instruction.
Kangaroo OMNI™ Item Codes Depending on regional availability, the following item numbers can be used for ordering Kangaroo OMNI™ Enteral Feeding Pumps, feeding sets, or accessory components: DESCRIPTION TYPE STERILE B5FD Kangaroo OMNI™ Feeding Set 500 mL E5FD Kangaroo OMNI™ ENtelliSet™ Feeding Set 500 mL B10FD Kangaroo OMNI™...
Page 72
Kangaroo OMNI™ Item Codes DESCRIPTION TYPE STERILE E10FDR Kangaroo OMNI™ ENtelliSet™ Thick Formula Feeding Set 1000 mL E10FFR Kangaroo OMNI™ ENtelliSet™ Thick Formula Feeding Set with Flush Bag 1000 mL E5FDSR Kangaroo OMNI™ ENtelliSet™ Thick Formula Feeding Set 500 mL Sterile E10FDSR Kangaroo OMNI™...
Cardinal Health. This Limited Warranty as applied to Kangaroo OMNI™ Enteral Feeding Pump battery and power adapter is limited to one (1) year from the date of shipment from Cardinal Health. 2. This Limited Warranty does not extend to routine maintenance of Kangaroo OMNI™ Enteral Feeding Pump.
Electromagnetic Conformity Declaration Guidance and Manufacturer’s Declaration – Electromagnetic Emissions Kangaroo OMNI™ Enteral Feeding Pump are intended for use in the electromagnetic environment specified below. The customer or the user of Kangaroo OMNI™ Enteral Feeding Pump should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment - Guidance...
Page 75
Electromagnetic Conformity Declaration Guidance and Manufacturer’s Declaration – Electromagnetic Immunity Kangaroo OMNI™ Enteral Feeding Pump are intended for use in the electromagnetic environment specified below. The customer or the user of Kangaroo OMNI™ Enteral Feeding Pump should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment -Guidance...
Page 76
Electromagnetic Conformity Declaration Guidance and Manufacturer’s Declaration – Electromagnetic Immunity Kangaroo OMNI™ Enteral Feeding Pump are intended for use in the electromagnetic environment specified below. The customer or the user of Kangaroo OMNI™ Enteral Feeding Pump should assure that it is used in such an environment. Immunity Test IEC 60601 test level Compliance...
Electromagnetic Conformity Declaration Recommended separation distances between portable and mobile RF communications equipment and the Kangaroo OMNI™ Enteral Feeding Pump Kangaroo OMNI™ Enteral Feeding Pump are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of Kangaroo OMNI™...
Appendix Appendix A - Accuracy Graphs The following graphs illustrate the accuracy of Kangaroo OMNI™ Enteral Feeding Pump and 2nd and 23rd hours, per the IEC 60601-2-24 standard. The graphs are shown for both the Minimum Rate (1 mL/hr) and the Intermediate Rate (50 mL/hr).
Page 79
Appendix A - Accuracy Graphs 23rd Hour Curve on Accuracy...
Page 81
Appendix A - Accuracy Graphs Ambient Temperature Effect on Accuracy The diagrams below show how ambient temperature affects accuracy. Ambient Temperature Effect on Accuracy of Water and Ambient Temperature Effect on Accuracy of Water and Low Low Density Formulas (<1.07 g/mL) Density Formulas (<1.07 g/mL) 4.0% 5.0%...
Page 82
Appendix A - Accuracy Graphs Head Height Effect on Accuracy The diagram below show how variation from the recommend hang height affects accuracy. Effects of Back Pressure on Accuracy The nominal back pressure Kangaroo OMNI™ Enteral Feeding Pump accuracy testing is established at the distal connector on the end of the feeding set.
Appendix B - Explanation of Alarms The purpose of this appendix is to provide an overview of how the Kangaroo OMNI™ Enteral Feeding Pump alarm system works and what priorities are assigned to each alarm. Overview Kangaroo OMNI™ Enteral Feeding Pump has alarms that are broken into three different priorities: High Priority (Red Blinking Indicator Light), Medium Priority (Yellow Blinking Indicator Light), and Low Priority (Solid Yellow Indicator Light).
Page 84
Appendix B - Explanation of Alarms Cassette Dislodged Alarm • Sensor Input: Magnetic Sensor Kangaroo OMNI™ Enteral Feeding Pump will alarm for Cassette Dislodged when a low voltage level is received from the magnetic sensor. Valve Error • Sensor Input: Motor Encoder Signal Kangaroo OMNI™...
Page 85
Appendix B - Explanation of Alarms Priority Handling of Information Signals In all cases, High Priority alarms are the most important and override any other alarm conditions. A medium or low priority alarm will never disable a High Priority alarm. Medium Priority alarms all have equal weighting. There should never be a situation when medium priority alarms are occurring at the same time, so there is no need to assign a weighting within the medium alarm priority.
Need help?
Do you have a question about the Kangaroo Omni 385400 and is the answer not in the manual?
Questions and answers
Why does the alarm for blockage on patient side keep going off even though there is no blockage. I have changed the set,I have cleaned the machine and every 5 minutes it is going off again
The blockage alarm on the Cardinal Health Kangaroo Omni 385400 could be caused by an issue with the occlusion sensor region being dirty or malfunctioning. To address this, remove the cassette, clean and dry the occlusion sensor region, and reload the cassette. If the alarm continues, the sensor may still be detecting an issue despite no actual blockage.
This answer is automatically generated
Mine does the same thing, the machine isn't even 2 weeks old. You can leave it running with the end unattached and see it's running fine and sometimes it just keeps sounding the alarm every couple of minutes. You can even **** on it like a straw so the sensor should be reading it put out faster than intended, still says it (lol). I increased the flush frquency to every hour and reduced the amount proportionally and it happens much less often now. Just for reference mine is being loaded with Kate Farms 1.5 peptide.
How to change feed time on kangaroo omni
To change the feed time on the Cardinal Health Kangaroo Omni 385400, follow these steps:
1. Go to the appropriate feed setup screen.
2. Use the controls to adjust the time interval for feeding.
3. Press the OK button to confirm the time interval and return to the Adjust Flush screen.
4. Press the Done button to confirm and return to the Ready to Feed screen.
Adjustments can only be made before starting the feed. If feeding has already started, press the Stop button to make changes.
This answer is automatically generated
How to clear volume on kangaroo Omni pump
To clear volume on the Cardinal Health Kangaroo Omni 385400 pump:
1. Go to the More Options screen.
2. Use the arrow keys to select "Clear Total Volumes" and press the Select button.
3. Press the Clear button.
4. Press the Back button to return to More Options.
This answer is automatically generated
When using Auto Prime the formula exits the cassette filled with air bubbles. Removed cassette and cleaned area. No resolution of problems. No alarms
Air bubbles in the cassette when using Auto Prime with the Cardinal Health Omni 385400 could be caused by air trapped in the bag or tubing before priming. This may happen if the feeding bag is not properly filled or if the tubing is not fully primed, allowing air to enter the system.
This answer is automatically generated
After priming the screen with start is not appearing
The start screen may not be appearing after priming the Cardinal Health Kangaroo Omni 385400 due to an incomplete priming process or an issue with the pump setup. Ensure that:
1. The feeding solution has exited the distal connector before pressing "Done."
2. The patient is disconnected during priming.
3. The cassette is properly loaded in the pump.
4. There are no alarms, such as "Feed Bag Empty" or "Patient Tube Blocked."
If the issue persists, try reloading the cassette and ensuring all steps are completed correctly.
This answer is automatically generated
How do I set language back to english
I am wondering more about the wifi connection possibilities
can the pump be set to flush 35 ml every hour