Changing Syringe and Restoring Infusion (Continued)
2.
To restart infusion using restored parameters, press
soft key and continue with next step.
RESTORE
OR
To start a new infusion, select drug from Drug Library and
follow steps for "Infusion Modes".
3.
Verify restored drug/concentration. Press
4.
Prime administration set (reference "Preparing Infusion",
"Priming").
5.
For restored parameters, verify parameters are valid and
press
CONFIRM
6.
Close and lock security door.
7.
Verify programming parameters on summary screen are
correct and press
NOTES:
If drug and/or drug concentration is different from previous
syringe, attach and prime new administration set.
To change a restored parameter:
a. Press applicable soft key.
b. Enter desired parameter using numeric data entry keys.
c. Press
CONFIRM
3-30
Programming
Infusion Modes (Continued)
soft key.
soft key.
START
soft key.
soft key.
NEXT
Patient Controlled Analgesia Module Section
A
Guardrails Drug Setup
Med Surg
Morphine
Meperidine
>Select Drug
EXIT
RESTORE
A
Second Nurse Summary
Morphine
2
PCA Dose:
mg
Lockout
10
Interval:
25
Max Limit:
[Conc]: 1 mg/mL
>Press START
PRO-
PAUSE
GRAM
Alaris
System Directions for Use
®
A-E
F-J
K-O
P-T
U-Z
minutes
mg/4 h
START