Cardinal Health Alaris 8000 Series Directions For Use Manual page 60

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Preparing Administration Set (
Loading (Continued)
c. Press safety clamp fitment into recess below
mechanism.
d. Using a finger tip, firmly push tubing toward back of
Air-in-Line (
AIL
4.
Close door and latch, as follows:
a. Close door and hold in a closed position
by grasping door and instrument case
with one hand.
b. Gently lower latch.
• Safety clamp device is automatically
disengaged.
5.
Open roller clamp.
6.
Verify no fluid is flowing through drip chamber.
2-4
Getting Started
) Detector.
) (Continued)
Pump Module
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
CAUTION
To reduce the potential for nuisance
alarms, ensure tubing is fully
AIL
inserted in
Detector.
AIL
Safety Clamp Fitment
Push tubing toward
back of AIL Detector.
WARNINGS
• Do not touch the administration
set while closing the door.
Failure to follow this instruction
may result in infusion rate
inaccuracy.
• To prevent a potential free-flow
condition, ensure no extraneous
object (for example, bedding,
tubing, glove) is enclosed or
caught in the Pump Module door.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section