Power Down All Channels

1.
Press
OPTIONS
2.
Press Power Down All Channels soft key.
3.
Press Yes soft key.
• During power off sequence, Main Display flashes
POWERING DOWN
When the Anesthesia Mode is enabled while a module is
paused, the module remains in an indefinite pause until
restarted.
When Anesthesia Mode is enabled:
• All limits are set to Soft.
• Dose checking mode is set to Smart.
• Key-press audio is turned off.
• Tamper Resist Mode (panel locked) is not available.
• All Guardrails
Drug entries are available for selection.
®
-- Continued Next Page --
1-16
General Setup and Operation
System Options (Continued)
key.
.
Anesthesia Mode
System Options 1 of 2
Display Contrast
Patient ID
Time of Day
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
System Options
Power Down
All Channels?
>Press Yes or No
CAUTION
When the Alaris
System is set up
®
for use in Anesthesia Mode, it is
important to select the profile that
corresponds with the care area the
patient will be taken to when the
Anesthesia Mode is discontinued.
This ensures that the Alaris
will be in the correct profile following
the use of the Anesthesia Mode.
Alaris
System Directions for Use
®
PC Unit Section
PAGE
DOWN
Yes
No
System
®