Table 3 – Parameter Default Value
Mode
NCPAP
Parameter
T-High
Default for NCPAP apply
Rate
Rb
T
/ T
This setting applies to all modes: system-wide default applies to all modes
apnea
LBR
User Interface Display
Screen Displays
1.
Set Up Screen – The Set Up screen prompts the user to confirm settings for
base line pressure level (NCPAP / Pres Low), %O
High) controls and confirmation of connection of the Transducer Interface (XDCR) to
the driver.
Adjust the NCPAP / Pres Low flow meter counter clockwise to increase the control to
the required flow rate and clockwise to decrease the flow. Touch the associated
flashing icon to confirm. The icon changes to a check mark and the next button
flashes.
Set the %O
as prescribed. Touch the associated flashing button to confirm. The
2
icon changes to a check mark and the next button flashes.
Adjust the Pres High flow meter as desired for delivery of BiPhasic, BiPhasic tr or
manual breaths using counterclockwise turns to increase and clockwise turns to
decrease the flow. Touch the associated flashing button to confirm. The icon
changes to a check mark and the next button flashes.
675-101-101
Revision H
NCPAP + rate
BiPhasic
monitoring
Default for BiPhasic apply
Default for BiPhasic apply
Figure 8 – Set Up Screen
BiPhasic + rate
monitoring
, upper level pressure (Pres
2
27
BiPhasic tr
Default for
BiPhasic tr
Default for
BiPhasic tr