Electromagnetic Environment (Continued)
Alaris
System (Continued)
®
Immunity Test
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
Alaris
System Directions for Use
®
Appendix Section
Compliance (Continued)
Electromagnetic Immunity - Life Support Equipment
IEC 60601-1-2
Compliance
Test Level
3 Vrms
10 Vrms
150 kHz - 80 MHz
10 V/m
3 V/m
80 MHz - 2.5 GHz
Table 3
Electromagnetic Environment - Guidance
Level
Portable and mobile RF communications equipment
should be used no closer to Alaris
(including cables) than recommended separation
distance calculated from equation applicable to
frequency of transmitter.
Recommended Separation Distance
12
— d = [-----] √P
V
12
— d = [-----] √P 80 MHz - 800 MHz
E
12
— d = [-----] √P 80 MHz - 2.5 GHz
E
d = recommended separation distance in
meters (m).
P = maximum output power rating of transmitter in
watts (W) according to transmitter manufacturer.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than compliance level in each
frequency range.
Interference may occur in vicinity of equipment
marked with following symbol:
2
1
1
Regulations and Standards
System
®
A-9