Secondary Infusion
6.
Using hanger provided with secondary administration set,
lower primary fluid container until bottom of secondary
container is at least
container.
Hanger
at least 9½"
Primary Container
The following procedure should be used only when:
• drug to be infused is listed in Guardrails
• primary infusion is running, and
• a check valve administration set is being used.
To program a primary infusion, reference "IV Fluid Infusion"
procedure. To program a basic infusion, reference "Infusion -
Guardrails
Suite MX Protection" procedure.
NO
®
1.
Press
CHANNEL SELECT
2-40
Programming
- With Guardrails
(
Pump Module
Setup (Continued)
" above fluid level in primary
9½
Infusion
key.
Suite MX Protection
®
) (Continued)
Secondary Container
Drug Library,
®
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Alaris
System Directions for Use
®
Pump and Syringe Modules Section