Suite Mx Protection (Pump Module ); Introduction And Setup; Infusion - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Secondary Infusion - NO

Introduction and Setup

Reference "Secondary Infusion - With Guardrails
Protection".
The following procedure should be used only when:
• drug to be infused is not listed in Drug Library,
• primary infusion is running, and
• a check valve administration set is being used.
To program a primary infusion, reference "Primary Infusion -
With Guardrails
Suite MX Protection", "IV Fluid Infusion"
®
procedure. To program a basic infusion, reference "Infusion -
Guardrails
Suite MX Protection".
NO
®
1.
Press
SECONDARY
Continue with next step.
2.
Enter secondary infusion rate or duration, as follows:
• To enter secondary infusion rate, press
and use numeric data entry keys.
• To enter duration, press
numeric data entry keys.
3.
To enter secondary volume to be infused, press
key and use numeric data entry keys.
4.
Open clamp on secondary administration set.
5.
Verify correct infusion parameters and press
key.
Changing Primary Infusion Parameter
1.
Press
CHANNEL SELECT
2.
Press
PRIMARY
3.
To change primary infusion parameter, press applicable
soft key (
or
RATE
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
(
Pump Module

Infusion

soft key and then
DURATION
key.
soft key.
), and use numeric data entry keys.
VTBI
Guardrails
Suite MX Protection
®
)
Suite MX
®
soft key.
BASIC SEC
soft key
RATE
soft key and use
soft
VTBI
soft
START
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
2-51
Programming