Attaching and Detaching Modules (Continued)
1.
Ensure module(s) is powered off before detaching.
2.
Push module release latch and then rotate module(s) up
and away from PC Unit or attached module (opposite to
motion shown above) to disengage connectors.
• Alaris
System reidentifies and shows appropriate
®
module identification (
• Appropriate module position(s) (
remaining module(s) appear on Main Display.
Adding Module(s) While System is Powered On
Add module as described in "Attaching Module(s)".
• System tests module, causing all
indicator lights of displays to illuminate briefly.
• Appropriate module identification display (
illuminates. Modules are always labeled left to right, so if a
module is added to left of other modules, all modules will
be reidentified. Module reidentification does
or affect infusion or monitoring on active modules.
• Module positions (
NOTE:
If any of the following conditions are observed, the affected
module must be removed from use and inspected by qualified
personnel:
segments are not illuminated on displays during
LED
power-on test.
Indicator lights do not illuminate.
Appropriate module identification (
displayed.
If the affected module operates normally when it is attached
via the alternate IUI connector, it may be used until a
replacement module can be substituted.
1-4
General Setup and Operation
Detaching Module(s)
,
,
or
A
B
C
D
A
LED
,
,
or
) appear on Main Display.
A
B
C
D
A
), from left to right.
,
or
) for
B
C
segments and
,
,
or
)
A
B
C
D
interrupt
NOT
,
,
or
) is not
B
C
D
Alaris
System Directions for Use
®
PC Unit Section

Advertisement

Table of Contents
loading

Table of Contents