Electromagnetic Environment (Continued)
Alaris
System (Continued)
®
Emissions Test
IEC 61000-4-2
Electrostatic
Discharge (
)
ESD
IEC 61000-4-4
Electrical Fast
Transient, Burst
(EFT)
IEC 61000-4-5
Power Line Surge
IEC 61000-4-8
Power Frequency
Magnetic Field
(50/60 Hz)
Alaris
System Directions for Use
®
Appendix Section
Compliance (Continued)
Electromagnetic Immunity
IEC 60601-1-2
Compliance
Test Level
±6 kV contact
±8 kV contact
±8 kV air
±15 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for
±1 kV for
input/output lines
input/output lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common
±2 kV common
mode
mode
3 A/m
400 A/m 50 Hz
400 A/m 60 Hz
-- Continued Next Page --
Table 2
Electromagnetic Environment - Guidance
Level
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material,
relative humidity should be at least
If connector testing exemption is used, following
sensitivity symbol appears adjacent to each
ESD
connector.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
.
30%
" - Do Not Touch"
Regulations and Standards
A-7