The Alaris
®
System Technical Service Manual is available
from Cardinal Health. It includes routine service schedules,
interconnect diagrams, component parts lists and descriptions,
test procedures, and other technical information, to assist
qualified service personnel in repair and maintenance of the
instrument's repairable components. Maintenance procedures
are intended to be performed only by qualified personnel,
using the Service Manual and System Maintenance software.
To enhance safety and ease of operation, the Alaris
provides a full range of audio and visual alarms, errors, and
messages.
Operating the system near equipment which radiates high-
energy radio frequencies (such as, electrosurgical/cauterizing
equipment, portable radios, cellular telephones) may cause
false alarm conditions. If this happens, reposition the Alaris
System away from the source of interference or turn off the
system and manually regulate the flow with the clamp and/or
monitor the vital parameters using an appropriate clinical
alternative
See the product-specific Section of this DFU that applies to the attached module(s) for alarm, error and
message definitions specific to that module.
Advisory / Message
Alarm
Alarm Silence
Alaris
®
System (with v9 Model 8000) DFU
PC Unit Section

General

Alarms, Errors, Messages

Definitions

A sequence of audio and/or visual signals indicating operating status of
Alaris
®
System.
An audio and visual signal that a potentially unsafe condition is
present. Immediate action is required.
Alarms can be silenced for up to
key. Alarm indicators remain on and, if applicable, alarm silence
symbol (see monitoring module Section of this DFU) is displayed.
Silence period can be ended by pressing

Troubleshooting and Maintenance

®
System
®
seconds by pressing
120
Troubleshooting and Maintenance
SILENCE
soft key.
CANCEL SILENCE