Measurement Support Functions - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B

Measurement Support Functions

Display
2
40-digits
2-lines LCD display. Capable of displaying:
measurement values, comparator/contact check decisions, control settings, self test messages,
and annunciators, annunciator labels.
Open Correction
Eliminates measurement errors due to the leakage in the test cable/test xture.
Comparator
HIGH/IN/LOW for the measurement parameter of each test channel
Contact Check
Contact failure between the test xture and device can be detected.
GPIB Interface
All control settings, measurement values, self-test results, and comparator information.
Handler Interface
All outputs signals are negative logic, opto-isolated open collector outputs
Output Signals Include:
HIGH/IN/LOW and no contact for each channel, index, end-of-measurement, and alarm
Input Signals Include:
Keylock and external trigger
Save/Recall
Ten instrument setups can be saved/recalled from the internal non-volatile memory.
Continuous Memory Capability
If the instrument is turned OFF, or if a power failure occurs, the instrument settings are
automatically saved.
Key Lock
Disables key input from the front panel.
General Information
8-5