Table 2- 6
average mode when in power or frequency sweep.

Table 2-6 Power and frequency sweep measurement functions table

Refer to the P-Series Power Meters Programming Guide for the commands usage details.
N O T E
Example of command set:
SENSe:BUFFer:MTYPe "PEAK"
Power Sweep Mode
Power sweep is generally used in power level calibration setup where the
frequency is fixed, and the amplitude of the power source signal is swept.
This mode can be used to characterize the flatness, linearity or gain
compression of a device under test.
This feature is applicable for 8480 Series, N8480 Series, E-Series E4410, E-Series E9300,
N O T E
E-Series E9320 or P-Series N1920 sensor when connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a
N1911A/1912A P-Series Power Meters User's Guide
shows the available measurement functions during normal and
Measurement functions
Hold off
Trigger delay
Measurement averaging
Gate setting
Channel offset
Hysteresis
Measurement types
supported
BNC cable. The same connection is applied for power source TRIG OUT
to power meter TRIG IN. See
General Power Meter Functions
Normal Mode
Peak, peak-to-average,
average, minimun power
Figure 2- 58
Average Mode
Average only
on page 84.