E4990A Help
Analyzer.WriteString ":SENS1:ADAP:CORR:COLL:SAVE PHAS", True
MsgBox "Phase Setup Done"
' Impedance setup
MsgBox "Connect Open Termination"
' Execute open in impedance setup
Analyzer.WriteString ":SENS1:ADAP:CORR:COLL:ACQ OPEN", True
' Wait for measurement end
Analyzer.WriteString "*OPC?", True
Dmy = Analyzer.ReadNumber
MsgBox "Connect Short Termination"
' Execute short in impedance setup
Analyzer.WriteString ":SENS1:ADAP:CORR:COLL:ACQ SHOR", True
' Wait for measurement end
Analyzer.WriteString "*OPC?", True
Dmy = Analyzer.ReadNumber
MsgBox "Connect LOAD Termination"
' Execute load in impedance setup
Analyzer.WriteString ":SENS1:ADAP:CORR:COLL:ACQ LOAD", True
' Wait for measurement end
Analyzer.WriteString "*OPC?", True
Dmy = Analyzer.ReadNumber
' Save impedance setup data
Analyzer.WriteString ":SENS1:ADAP:CORR:COLL:SAVE IMP", True
MsgBox "Imepdance Setup Done"
' Close IO
Analyzer.IO.Close
End Sub
Preparing for Accurate Measurement
Preparing for Accurate Measurement
User Calibration
Fixture Compensation
User Calibration
260