Setting the Measurement Display

Using the gate settings, up to four measurements or a combination of
measurements can be shown on the display. Any of the 4 measurement display
lines can show any of the measurement results from any of the 4 gates, allowing
you complete control of the displayed information.
The power meter cannot display all 16 (or 32) measurements simultaneously.
Average, peak, peak-to-average ratio, and minimum measurements are made
within each gate period generating possible measurement results per channel as
shown in
Figure 2-52
122
Figure
2-52.
Gates 1 to 4
Peak
Average
Gate 1
Pk-to-Avg
Min
Peak
Average
Gate 2
Pk-to-Avg
Min
Peak
Average
Gate 3
Pk-to-Avg
Min
Peak
Average
Gate 4
Pk-to-Avg
Min
Sixteen measurements per channel
Feed 1
Feed 2
Feed 1
Feed 2
Feed 1
Feed 2
Feed 1
Feed 2
Keysight N1911A/1912A User's Guide