Minimum temperatures are also affected by the ambient humidity. If the
humidity is high, the minimum temperatures achieved for a temperature zone
is not as low as those achieved in a low ambient humidity environment.
NOTE
A DC Temperature Calibration Kit (P/N 063782) is available.
The kit includes the parts and instructions required to verify
the temperature calibration of the DC upper and lower com-
partments and to recalibrate, if required.
To set the fan speed of the DC upper compartment to high:
1. Open the Chromeleon 7 Instrument Configuration program or
Chromeleon 6.8 Server Configuration program.
2. Double-click the DC icon under the instrument or timebase.
3. Click the Thermal Controls tab and then double-click
Compartment_TC.
4. Under Fan Speed, click High Speed (see
Figure 2-23. Setting the DC Upper Compartment Fan Speed
64
Figure
2-23).
Doc. 065342-06 12/11