When combining aqueous and nonaqueous components (for example,
water and acetonitrile). Pressurizable reservoirs allow eluents to be stored
under a specific atmosphere.
Never pressurize eluent reservoirs above 0.07 MPa (10 psi).
Pressurizing reservoirs above this limit can cause the reservoir to
explode.
Ne mettez jamais les réservoirs d'éluants sous une pression supérieure
à 0,07 MPa (10 psi).
Setzen Sie den Eluentbehälter auf keinen Fall einem Druck über
0,07 MPa aus.
If you plan to pressurize the eluent reservoirs, an optional regulator kit is
required. The kit is available in two versions:
When the DC is installed as the topmost module in the system, the EO
Regulator Kit (P/N 074422) is required. The kit includes a pressure
regulator and gauge assembly with four outputs (for connections to four
eluent reservoirs), as well as the tubing and connection fitting required.
When the TC, VWD, or PDA is installed as the topmost module in the
system, the TC/VWD/PDA Regulator Bracket Kit (P/N 074424) is
required. The kit includes the EO Regulator Kit described above, as well
as a right-angle regulator bracket and mounting hardware. After attaching
the bracket to the TC or detector, you will mount the gas regulator
assembly on the bracket.
If more reservoirs are required, order a second regulator (P/N 074423).
For instructions on installing the regulator kits, refer to Thermo Scientific
Dionex ICS-5000 Ion Chromatography System Installation Instructions
(Document No. 065343). The manual is included on the Thermo Scientific
Reference Library DVD (P/N 053891).
Doc. 065342-06 12/11