Error during storing of the card data to the chip card.
This error occurs if the column ID chip card is installed incorrectly, or is
defective.
To troubleshoot:
When the chip card is properly installed, the Dionex logo on the card faces up
and the LED on the memory slot is green. If this error occurs when the chip
card is properly installed, it indicates that the card is defective and should be
replaced (for instructions, refer to
Error programming flash memory.
This error occurs if, during a firmware download, the new firmware cannot be
programmed into memory.
To troubleshoot:
Flash memory cannot be serviced by the user. Contact Technical Support for
Dionex products for assistance.
Executing this command will reset associated counter values
stored in the module.
This error occurs when you select a command (RotorSealChanged, for
example) that updates information stored in the TC. To prevent information
from being inadvertently changed, Chromeleon requests confirmation before
executing the command.
To troubleshoot:
If you want to execute this command, click Execute; otherwise, click Cancel.