Page 3
28199 Bremen Germany Send an e-mail message to the Technical Editor at • documentation.bremen@thermofisher.com You are encouraged to report errors or omissions in the text or index. Thank you. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 4
Contacting Us Suggestions to the Manual iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Vacuum System .............. 3-11 Forepump ..............3-12 Turbomolecular Pump..........3-12 Pressure Measurement..........3-12 Gas Supply..............3-13 Connections..............3-13 Plasma and Cooling Gas ..........3-14 CCT Gas ..............3-15 Additional Gas ............. 3-17 Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 6
System Controls..............6-4 System Status LEDs ............6-4 Power Panel ..............6-5 Starting Up the System ............. 6-6 Shutting Down the System ..........6-7 Switching to Standby Mode ........... 6-7 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 7
Cleaning the Surface of the Instrument ......8-7 Cleaning Agents ............. 8-7 Checking Warning Labels ..........8-7 Checking Gas Lines for Leaks ........8-8 Maintaining the Air Filter ..........8-9 Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
It describes the modes of operation and principle hardware components of your iCAP Q mass spectrometer and provides step-by-step instructions for cleaning and maintaining your instrument. Thermo Fisher Scientific...
“press <Shift> + <F1>” means press and hold the <Shift> key and then press the <F1> key. Any button that you click on the screen is represented in bold face • letters. For example, “click Close”. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Topic Headings The following headings are used to show the organization of topics within a chapter: Chapter Name Chapter 1 Second Level Topics Third Level Topics Fourth Level Topics Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
The iCAP Q Operating Manual represents the Original Operating Instructions. Thermo Fisher Scientific provides additional reference documents for the iCAP Q mass spectrometer that are not part of the Original Operating Instructions. Reference documentation for the iCAP Q mass spectrometer includes the following: iCAP Q Preinstallation Requirements Guide •...
Equipment for connecting the basic instrument • Printed manuals iCAP Q Operating Manual For a list of the manuals delivered with the iCAP Q mass spectrometer and instructions on where to find PDF files of these manual, see “Reference Documentation” page 1-4.
System Description Chapter 3 This chapter provides an overview of the functional elements of the iCAP Q instrument. It contains the following topics: • “General Description” page 3-2 “Sample Introduction System” page 3-4 • “Ion Optics” page 3-9 • “QCell”...
System Description General Description General Description The iCAP Q mass spectrometer is designed for direct analyses of liquid samples and can be coupled to further devices for analyses of gases or solid aerosols. Three models are available: iCAP Qa, iCAP Qc and iCAP Qs.
System Description General Description Instrument Layout The iCAP Q instrument can be divided into four main components, which are described in the following topics: Sample introduction • Interface (ion generation) • Ion Optics (ion focussing) • Mass Analyzer • Labeled Components: 1= sample introduction, 2=interface, 3=ion optics, 4=mass analyzer Figure 3-2.
System Description Sample Introduction System Sample Introduction System The sample introduction system of the iCAP Q MS comprises the peristaltic pump, the nebulizer, the Peltier-cooled spray chamber and the torch with the injector. Figure 3-3 shows a schematic view of the sample introduction system.
Depending on the model ordered, the spray chamber is made of either quartz or PFA, see Table 3-1. Labeled Components: 1=nebulizer, 2=spray chamber, 3=injector, 4=torch Figure 3-5. Nebulizer, spray chamber, injector and torch Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
System Description Sample Introduction System For the iCAP Q instrument, the spray chamber is cooled using a Peltier cooled chiller to improve instrument performance. Injector The injector is available with various internal diameters (ID) and in various material for different types of analysis. Injectors with smaller...
The sample cone introduces ions from the plasma into the first vacuum stage (interface vacuum) and the skimmer cone admits the ions in the mass spectrometer. Figure 3-8. Sample cone and skimmer cone Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Figure 3-9. Detail of extraction lens The extraction lens focusses and accelerates the ions from the back of the skimmer cone into the intermediate vacuum region of the analyzer. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
The ion optics region comprises the RAPID lens, the QCell, and the DA assembly. RAPID Lens Ions extracted from the iCAP Q interface are accelerated to the RAPID (Right Angle Positive Ion Deflection) lens which deflects analyte ions by 90 ° before they enter the QCell.
RF voltage and DC voltage applied to the quadrupole rods. Detector Ions transmitted through the quadrupole are finally transferred to the dual mode secondary electron multiplier (SEM) and detected. 3-10 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
The ICP source operates at atmospheric pressure while the mass analyzer and detector requires a high vacuum for optimum performance. The iCAP Q instrument has a three-stage pumping system realized by a rotary pump and split-flow turbomolecular pump. Figure 3-11 shows a schematic view of the vacuum stage in the instrument.
The pressure values are read back to the Dashboard of the Experiment Editor, see Figure 3-12. The values are also read back in the Vacuum pane of the Instrument Control window. Figure 3-12. Experiment Editor - Dashboard 3-12 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
System Description Gas Supply Gas Supply The iCAP Q MS uses argon gas to generate the inductively coupled plasma and to control internal functions with the aid of pneumatics. Additional gases may be required depending on the type of analysis.
Inside the iCAP Q instrument, tubes are already installed from the connector to the optional mass flow controller. The connectors for additional gases are also situated on the left side of the iCAP Q instrument, see 4 and 5 in Figure 3-13.
System Description Gas Supply Connection of the laboratory argon supply to the iCAP Q instrument is made via a push-in fitting. A recommended set-up for the argon supply is shown in Figure 3-15: Labeled Components: 1=pressure regulator, 2=shut-off valve, 3=pressure gauge, 4=argon inlet, 5=iCAP Q, 6=gas source Figure 3-15.
Page 30
Thermo Fisher Scientific does not endorse any manufacturer, nor does it endorse products other than its own. Companies and products listed in this guide are given as examples only. 3-16 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Connections for Water Supply The iCAP Q system can be supplied with cooling water either from a central system or from the optionally shipped recirculation water chiller NESLAB ThermoFlex 2500. The connections for the cooling water are...
System Description Cooling System Recirculating Chiller Optionally, the iCAP Q mass spectrometer is delivered with the recirculating chiller NESLAB ThermoFlex 2500 with closed circuit, cooled by a refrigerating device. NOTICE For more information, refer to the manufacturer’s manual supplied with the instrument and see “Recirculating Chiller”...
System Description Exhaust System Exhaust System An exhaust system is needed for the iCAP Q MS to remove gases that may contain ozone and other noxious substances. Two exhaust ports are situated at the top of the instrument, see Figure 3-17.
Connections The main power switch, the mains inlet, the water supply connections, the pumping connections, and the gas supply connections of the iCAP Q are located at the left side of the instrument, see Figure 3-18. Labeled Components: 1=main power switch, 2=main power supply, 3=water connections, 4=pumping connections, 5=gas connection Figure 3-18.
Figure 3-20. Main power switch of iCAP Q MS Power Connections The main power supply connection and the power connection for the fore vacuum pump are located at the left side of the iCAP Q mass spectrometer, see Figure 3-21.
System Description Connections Connection to Computer and Peripherals The peripheral control connections allows transmitting digital contact closure signals from the iCAP Q mass spectrometer to other systems such as autosamplers or LC systems. Figure 3-23 shows the connection panel at the right side of iCAP Q MS.
Page 40
System Description Connections 3-26 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Be sure to read and comply with all precautions described in this manual. System configuration and specifications in this manual supersede all previous information received by the purchaser. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
The label indicates the presence of electricity at the main power switch. The label indicates the presence of high voltage inside the instrument. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 44
The label indicates the presence of noxious gases at the plasma exhaust. The label marks the gas connections. Only permitted gases may be connected here, see “Permitted Materials” page 4-7. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
615 MODEL: iCAP Q SERIAL No: This device complies with Part 18 of the FCC Rules Figure 4-1. Rating plate at the rear side of the iCAP Q MS Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
• designed for use in diagnostic or medical therapeutic procedures. If the iCAP Q mass spectrometer is used in a manner not specified by Thermo Fisher Scientific, the protection provided by the instrument could be impaired. Thermo Fisher Scientific assumes no responsibility and will not be liable for instrument damage and/or operator injury.
Permitted Materials Qualified Personnel Only employees of Thermo Fisher Scientific or personnel acting on behalf of Thermo Fisher Scientific are permitted to install the iCAP Q mass spectrometer. Personnel that installs or operates iCAP Q mass spectrometer must have the following qualifications: Electrical Connections •...
Do not place any objects upon the instrument—especially not • containers with liquids—unless it is requested by the user documentation. Leaking liquids might get into contact with electronic components and cause a short circuit. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
By making the operator aware of the potential risks, we hope to ensure maximum safety in the operation of this mass spectrometer. Users of the iCAP Q mass spectrometer must pay attention to the following residual hazards. WARNING Toxic Gases.
19-120-2947C [size large]; Thermo Scientific P/N 23827-0008) [size medium] or P/N 23827-0009 [size large]). For handling hot objects, gloves made of heat-resistant materials (for example, leather) should be available. 4-10 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Notice on the Susceptibility to Electromagnetic Transmissions Your instrument is designed to work in a controlled electromagnetic environment. Do not use radio frequency transmitters, such as mobile phones, in close proximity to the instrument. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 4-11...
Figure 4-2 shows the plasma door handle: Labeled Components: 1=plasma door closed, 2=plasma door open Figure 4-2. Plasma door handle at front of iCAP Q MS ❖ To shut off the plasma in case of emergency 1. Turn the handle 1 in Figure 4-2 clockwise for 180 degrees.
Page 53
5-7 • Only employees of Thermo Fisher Scientific or personnel acting on behalf of Thermo Fisher Scientific are allowed to install the iCAP Q mass spectrometer. Only employees of Thermo Fisher Scientific or personnel acting on behalf of Thermo Fisher Scientific are allowed to install the additional gas module (MFC).
CAUTION Hazardous Gases. Danger to health. When operating the instrument with hazardous gas, an external warning sensor must be installed in the vicinity of the instrument. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
System Setup Placing the System Placing the System The iCAP Q instrument is a bench top instrument to be placed on top of a table with its rear panel against a wall. Instrument Dimensions Figure 5-1 shows the typical required floor space for the system without autosampler.
1400 700-1000 2100-2400 3500-3800 Labeled Components: 1=monitor, 2=mouse, 3=keyboard, 4=iCAP Q MS, 5=wall, 6=autosampler housing, 7= computer table Figure 5-2. Required floor space for iCAP Q MS with autosampler Moving the Instrument The iCAP Q mass spectrometer is provided with four retractable handles for transport.
(autosampler, liquid chromatograph, for example) and stand in a secure and level position. Only workbenches with four legs provide sufficient stability for the iCAP Q mass spectrometer. The workbench must be dry and clean. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 58
It is recommended to leave sufficient space on the right side of the iCAP Q system for the connections to the computer and peripheral devices. NOTICE Avoid blocking the ventilation slots at the side of the instrument.
16 A. See rating plate (page 4-5). The iCAP Q instrument must have a separate “clean” line leading to a main fuse to guarantee disturbance-free operation. Gas Supply The iCAP Q system requires argon gas to generate the inductively coupled plasma and for controlling internal functions with the aid of pneumatics.
▲ Altitude The iCAP Q instrument is designed for indoor use at an altitude of up to 2000 m (6562 ft) above sea level. For altitudes above 2000 m, contact Thermo Fisher Scientific.
▲ Airborne Noise Emission The mean value for the emission sound pressure level of the iCAP Q instrument is below 70 dB. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 62
System Setup Laboratory Conditions 5-10 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Operation Safety Guidelines for Operation Safety Guidelines for Operation When operating the iCAP Q system, please pay attention to the following general safety guidelines. WARNING Toxic Gases. Danger of intoxication. Toxic gases are released that might lead to severe bodily harm when the plasma exhaust is not connected to an exhaust system or the spray chamber is disassembled.
Page 65
CAUTION Rotating Parts Hazard. Danger of injuries. The peristaltic pump rollers can pull in hair, clothing, ties and other loose objects. Stay clear from the peristaltic pump during operation. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Figure 6-1. ICP-MS front with system control LEDs System Status LEDs Three LEDs at the front of the instrument (see Figure 6-1) indicate the status of the main items of the iCAP Q MS. Table 6-1 lists their meaning. Table 6-1.
Figure 6-2. iCAP Q MS front with main power switch detail In the OFF position, the power to the instrument and the forepump is switched off. In the ON position, power is supplied to the instrument.
4. If the system was vented, the detector should be calibrated regardless of whether it was changed or not. The Get Ready procedure can now be started, see “Getting Ready” page 6-9. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
NOTICE If the plasma is not lit, start with step ▲ ❖ To shut down the instrument 1. Click inside the Experiment Editor window. The plasma is switched off. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
▲ ❖ To shut down the instrument 1. Click inside the Experiment Editor window. The plasma is switched off. 2. Switch the main power switch to Off position. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Getting Ready Getting Ready Thermo Fisher Scientific provides a tool to easily start up the iCAP Q mass spectrometer. An automatic sequence can be started that switches on the plasma and will carry out a routine check of the instrument’s performance by generating a performance report.
Page 72
Figure 6-4. Experiment Editor - Example of Get Ready window 4. If required, configure a delay for warmup of the instrument. 6-10 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 73
If the sensitivity test has been failed, a source autotune procedure is triggered. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 6-11...
Page 74
Get Ready procedure indicates its successful completion. If other modes have been selected in the Get Ready procedure, they will be tested once the STD mode performance report has been passed successfully. 6-12 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
❖ To switch on the instrument 1. Open Instrument Control. 2. Click On in the iCAP Q ribbon. The start-up procedure is triggered and the plasma is lit. 3. Select the measurement mode from the drop down menu in the Measurement Mode ribbon tab.
In the Browse for LabBook dialog box, select a LabBook. c. Click OK to open the LabBook. 3. The LabBook opens in a new tab of the Experiment Editor. 6-14 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
The acquisition queue itself is started independently once all LabBooks to be acquired are submitted. Starting the Data Acquisition ❖ To start the acquisition queue Click in the Scheduler pane to start data acquisition. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 6-15...
Emptying the waste container • Run the System with Blanks After operation it is recommended to flush the iCAP Q mass spectrometer with appropriate blanks to such as diluted nitric acid or ultrapure water.This ensures that neither high amounts of analytes nor matrix stay inside the sample introduction system over a longer period of time.
Instrument Control window are required. Safety Guidelines for Troubleshooting When performing troubleshooting on the iCAP Q system, please pay attention to the following general safety guidelines. CAUTION Service by the customer must be performed by trained qualified personnel only and is restricted to servicing mechanical parts.
Page 81
Thermo Fisher Scientific personnel. To ensure that the instrument is free from all electric current, always disconnect the power cords of mass spectrometer and fore vacuum pump before attempting any type of maintenance. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
12. Make sure the plasma torch is dry. 13. Make sure the torch body is aligned correctly inside the torch holder, see “Maintaining the Torch” page 8-15. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 84
18. Make sure the Forward Power during ignition is 900 W or greater. 19. Make sure that the cooling water meets the conductivity requirements. If the problem remains, contact the Qtegra Administrator for further troubleshooting. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Make sure that the cones and sample cone gasket are in good condition. If none of the above shows any improvements on the signal, contact the Qtegra Administrator for further troubleshooting. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
14. Check that the tension on the peristaltic pump tubing is correct and that the pump tubing is in good condition. If none of the above shows any improvements on the signal, contact the Qtegra Administrator for further troubleshooting. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
13. Make sure that the pump speed of the Peristaltic pump is stable. 14. Make sure that the vacuum system is stable. If none of the above shows any improvements on the signal, contact the Qtegra Administrator for further troubleshooting. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
To reset the Low Water Flow interlock 1. Make sure that the water cooler is switched on. 2. Make sure that the water cooling lines are properly connected. 7-10 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
1. Check that the argon supply is switched on and can supply 6 bar. 2. Make sure that the gas lines are securely connected, free from kinks, and not trapped. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 7-11...
Also see the readback values in the Plasma pane, Figure 7-1 page 7-4. “Plasma not Lit” page 7-5 for more information on the troubleshooting procedure. 7-12 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Maintenance Chapter 8 This chapter describes routine maintenance procedures that must be performed to ensure optimum performance of the iCAP Q mass spectrometer: “Safety Guidelines for Maintenance” page 8-2 • “General Advice for Maintenance” page 8-4 • “Servicing Plan” page 8-5 •...
Maintenance Safety Guidelines for Maintenance Safety Guidelines for Maintenance When performing maintenance on the iCAP Q system, please pay attention to the following general safety guidelines. WARNING Toxic Gases. Danger of intoxication. Toxic gases are released that might lead to severe bodily harm when the plasma exhaust is not connected to an exhaust system or the spray chamber is disassembled.
Page 93
Thermo Fisher Scientific field service engineer. Only use fuses of the type and current rating specified. Do not use repaired fuses and do not short-circuit the fuse holder. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Maintenance General Advice for Maintenance General Advice for Maintenance When performing maintenance on the iCAP Q mass spectrometer, observe the following advice: Accurate results can be obtained only if the system is in good • condition and properly calibrated. •...
Symbols on the Instrument” page 4-3 for an overview of the warning labels on the instrument. Also check the warning labels on the fore vacuum pump and the peripherals. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
5. When you cannot find a leak in the gas line, we recommend calling a Thermo Fisher Scientific field service engineer to check for gas leaks within the instrument. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Slightly lift up the cover. b. Pull off the cover. Figure 8-2. Open fan cover 2. Visual inspect the filter for dirt or damage. 3. Close the cover again. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 100
8-2. 2. Carefully pull out the filter. Figure 8-3. Remove filter 3. Clean the filter with a vacuum cleaner. 4. Insert the filter again. 5. Close the cover. 8-10 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
To remove the nebulizer 1. Pull off the tubing. 2. Carefully turn the nebulizer counterclockwise. 3. Pull out the nebulizer. Labeled Components: 1=tubing, 2=nebulizer Figure 8-4. Inlet system with nebulizer Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 8-11...
Pinch Point Hazard. Danger of injuries. Keep hands clear from the peristaltic pump during operation. 2. Wipe the surface of the peristaltic pump rollers and tension arms with a dry cloth to remove any dust. 8-12 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Pull back the holder spring to release the spray chamber. 3. Carefully pull out the spray chamber and the elbow connector from the torch assembly by pulling out the elbow connector. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 8-13...
8-11. 2. Remove the spray chamber, see “Removing the Spray Chamber” page 8-13. 3. Carefully remove the injector and the injector holder. Figure 8-5. Injector with holder 8-14 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
4. Allow it to air dry completely. If any components are to be stored for some time before use, they should be placed in metal-free sealed plastic bags. 8-16 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
“Shutting Down the System” page 6-7. Tools The iCAP Q instrument is provided with a cone tool for easy disassembly of the cones. The cone tool is magnetic to hold the cones in place when removing them from the interface block. Figure 8-7 shows the cone tools with the cones arranged at their respective tool side.
Page 108
Figure 8-9. Sample cone and holder in cone tool d. Remove the sample cone and the cone holder from the tool. The sample cone can now be cleaned. 8-18 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 109
Inserting the Sample Cone ❖ To insert the sample cone 1. Make sure that the sample cone gasket is in good condition, see Figure 8-11. Replace it if required. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 8-19...
Align the two lugs in the cone tool in the outer edge of the cone surface. c. Turn the tool counterclockwise, it will be held in place magnetically. 8-20 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 111
Maintaining the Skimmer Cone Insert ❖ To replace the skimmer cone insert using the insert tool 1. Align the insert tool with the holes in the skimmer cone insert. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 8-21...
Page 112
Figure 8-15. Figure 8-15. Skimmer cone insert and tool. NOTICE Make sure to mount the insert with the correct side facing the tool, see Figure 8-15. ▲ 8-22 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
2. Open the interface door. 3. Turn the extraction lens assembly counterclockwise, see Figure 8-16. Figure 8-16. Turn the extraction lens 4. Pull out the extraction lens assembly, see Figure 8-17. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 8-23...
Page 114
• Make yourself aware of all safety procedures in case of spillage or exposure to skin or eyes before commencing the cleaning. The acid cleaning should take place in a fume hood. 8-24 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 115
Figure 8-17. Be careful not to tilt the assembly. 3. Turn the extraction lens clockwise to fix it to the extraction lens housing. Figure 8-19. Insert extraction lens Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B) 8-25...
The Preventive Maintenance Guide provided by Thermo Fisher Scientific lists services that must be performed by a Thermo Fisher Scientific field service engineer or their representatives only. Refer to this guide for more information. 8-26 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 117
Additional Safety Information Chapter 9 This chapter contains excerpts from component suppliers’ documents with safety-relevant information that may concern users of the iCAP Q mass spectrometer. It contains the following topics: “Fore Vacuum Pump” page 9-2 • “Recirculating Chiller” page 9-4 •...
ATEX areas. Contact the pump manufacturer if you are planning such an application. Before installing the pump, you must reliably disconnect it from the electrical power supply and prevent the pump form running up inadvertently. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 119
Do not smoke with PFPE on your fingers. Touch the inner sections of the pumps only while wearing clean gloves, and use clean tools; do the necessary work in clean and dry rooms. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Use caution when removing the assembly. Water-cooled units also have a fan with sharp blades, ensure that the unit is off before removing the assembly. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Wear protective eye wear and gloves when handling the cartridge. The power supply cord is used as the main disconnect device. Make sure the socket-outlet is located near the Detector/Chromatography Module (DC) and is easily accessible. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 122
Additional Safety Information Dionex ICS-5000 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 123
Legal Documents This chapter provides legal documents. It contains the following documentation: • “FCC Compliance Statement” page D-2 “WEEE Compliance” page D-3 • • “Declaration of Conformity” page D-4 Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Legal Documents FCC Compliance Statement FCC Compliance Statement This device complies with part 18 of the FCC rules. iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 125
Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at http://www.thermoscientific.com/rohsweee. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 126
Legal Documents Declaration of Conformity Declaration of Conformity iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
12u; unit of mass for expressing masses of atoms or atomic mass (atomic mass constant). molecules. DAC Abbr. for digital-to-analog converter; a device that converts data from digital to analog form. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 128
HR Abbr. for High Resolution m/z mass-to-charge ratio HV high voltage Hz hertz (cycles per second) OD outside diameter ICIS™ Interactive Chemical Information System PCB printed circuit board ID inside diameter iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 129
WEEE European Union Waste Electrical and Electronic Equipment Directive. Provides guidelines square fit regression type: the fit is performed with a for disposal of electronic waste. second order (quadratic) function. Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
DA lens 3-10 DA plate 3-10 Dashboard 3-12 injector dashboard in-line filter 3-19 deposition, of matrix installation test 3-16 diagnosing, potential problems Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 132
6-10 STD mode organic solvents 3-17 system control LED oxygen 3-17 System LED system status LED Peltier cooler performance report temperature peristaltic pump 8-12 coolant water 7-8–7-9 iCAP Q Operating Manual (P/N 1288090, Revision B) Thermo Fisher Scientific...
Page 133
Vacuum LED torch vibration total dissolved solids level 7-8–7-9 tQuant 6-14 transport pallet trQuant 6-14 wall receptacle turbomolecular pump 3-12 warning labels Water Valve Closed interlock 7-11 Thermo Fisher Scientific iCAP Q Operating Manual (P/N 1288090, Revision B)
Page 134