Dionex ICS-5000 Operator's Manual
Tubing/Fitting Size
and Type
10-32 fitting bolt
10-32 ferrule, double-
cone
Table 9-3. Tubing and Fittings for a Capillary IC EG Channel
Notes for capillary IC channels:
Capillary
IC
Do not cut the 0.062 mm (0.0025 in) ID blue PEEK tubing used for
capillary IC connections. This tubing must be cut at the factory using a
specialized precision cutter. This ensures good chromatographic results. If
you need to replace any tubing of this type, order the appropriate
precision cut tubing assembly listed in
When replacing capillary tubing, make sure the ferrule and fitting bolt are
at least 2 mm (0.1 in) from the end of the tubing before you insert the
tubing into the port. Do not position the ferrule and fitting bolt flush with
the end of the tubing. For detailed connection instructions, see
Section
For tightening requirements for the high-pressure fitting bolt
(P/N 074449) and ferrule (P/N 074373), see
For tightening requirements for the 10-32 fitting bolt (P/N 043275) and
ferrule (P/N 043276), refer to Installation of Dionex Liquid Line Fittings
(Document No. 031432). The manual is on the Thermo Scientific
Reference Library DVD (P/N 053891).
334
Color
Black
9.1.
P/N
Use to connect...
043275
All other tubing connections
043276
Table
9-3.
Section
9.2.
Doc. 065342-06 12/11