The analog pressure output defaults to the pump 1 (bottom pump) output
signal. To redirect the output to pump 2 (the top pump), follow these steps:
1. Open the Chromeleon 7 ePanel Set or the Chromeleon 6.8 panel tabset.
2. Press the F8 key to open the Commands dialog box.
3. Select the pump name.
4. Select the AnalogOut property and select PumpTop. For Chromeleon
6.8, click Execute.
USB Connections
One USB receptacle ("B" type connector) allows a connection from the
PC on which Chromeleon is installed.
Three USB (Universal Serial Bus) ports ("A" type connectors) are
provided for connections to other Dionex ICS-5000 modules.
One 1.8 m (6 ft) USB cable (P/N 960777) is provided in the pump ship kit:
Pump Ship Kit
DP Ship Kit Dual Capillary
DP Ship Kit Dual Analytical
DP Ship Kit Capillary and Analytical
SP Ship Kit Capillary or Analytical
Waste and Drain Lines
One waste line is connected to the secondary pump head on each pump
and is routed through the tubing chace to the rear of the DP/SP.
The drain port removes any liquid that collects in the drip tray in the
bottom of the DP/SP. A clear corrugated drain hose (P/N 055075) is
connected to this port during installation.
Place the free ends of the waste lines and drain hose into a waste container. To
maintain a positive siphon, position the waste container below the level of the
pump.
For correct drainage, make sure the waste lines and drain hose are not
bent, pinched, or elevated at any point. Do not allow the ends of the
lines to be submerged in waste liquid.
Doc. 065342-06 12/11
2 • Description
Part Number
072112
062463
072111
063342
33