Thermo Scientific Dionex ICS-5000 Operator's Manual page 370

Ion chromatography system
Table of Contents

Advertisement

Dionex ICS-5000 Operator's Manual
To Gas Separator
WASTE
SEPARATOR
2. Check that no liquid lines are caught under the CR-TC. Align the two
mounting holes on the CR-TC back plate with the ball studs on the
mounting plate, and then push the CR-TC firmly onto the mounting ball
studs. The CR-TC clicks into place when properly installed.
3. Push the CR-TC electrical cable firmly into the
bulkhead connector. Twist the ring on the cable connector fingertight to
secure it.
4. Push the tray into the module until it locks into place. Close the EG door.
354
To
EGC
OUTLET Port
Waste Tube
,
GAS
Figure 9-29. Analytical CR-TC Plumbing Diagram:
Installation Completed
To RFIC Eluent
Degasser
REGEN OUT Port
To CR-TC
Regen In
(orange label)
TO CR
TC
-
REGEN OUT
(blue label)
To CR-TC
Eluent In
(red label)
CR-TC
CR-TC 1
To RFIC Eluent
Degasser
ELUENT IN Port
To CR-TC
Eluent Out
(yellow label)
CR-TC 2
(or
)
Doc. 065342-06 12/11