Thermo Scientific Dionex ICS-5000 Operator's Manual page 301

Ion chromatography system
Table of Contents

Advertisement

Firmware download in progress. Please wait.
This error occurs if you attempt to issue a command while firmware is being
downloaded to the TC.
To troubleshoot:
Wait until the firmware download is complete. Audit Trail messages will
inform you of the status of the download.
Gas leak detected.
This error occurs if the gas sensor detects an increased concentration of gas
inside the TC.
To troubleshoot:
1. Find and eliminate the leak.
2. If you are using Chromeleon 7: Open the Commands dialog box (press
the F8 key). Select the TC and click the Commands tab. Click the
ClearAlarmLED button.
3. If you are using Chromeleon 6.8: On the TC Control panel, click the
Clear Alarm button.
4. Ventilate the interior of the TC thoroughly before closing the door and
resuming operation.
Humidity leak detected.
This error occurs if the humidity sensor detects an increased concentration of
humidity inside the TC.
To troubleshoot:
1. Find and eliminate the leak.
2. If you are using Chromeleon 7: Open the Commands dialog box (press
the F8 key). Select the TC and click the Commands tab. Click the
ClearAlarmLED button.
Doc. 065342-06 12/11
8 • TC Troubleshooting