EG1 invalid flow.
-or-
EG2 invalid flow.
This error occurs if the flow rate is set to a value the EG does not support.
To troubleshoot:
Set the flow rate to a value within the allowed range. Refer to
flow rate specifications.
EG1 invalid flow rate-concentration.
-or-
EG2 invalid flow rate-concentration.
This error occurs if the selected concentration is too high for the current flow
rate.
To troubleshoot:
Set the flow rate to a value within the allowed range. The allowable eluent
concentration for a particular application depends on several factors: the flow
rate, suppressor type, EGC type, and cartridge configuration. For details, refer
to
Section
4.9.
EG1 invalid ion count.
-or-
EG2 invalid ion count.
This error occurs if the ion count reported by the EGC is invalid. This may
indicate a problem with the memory chip in the EGC.
To troubleshoot:
1. Make sure the EGC cable is securely plugged into the EG (see
Figure
2. If the error message appears again, contact Technical Support for Dionex
products for assistance.
246
2-10) and the locking ring is tightened.
Appendix A
for
Doc. 065342-06 12/11